Ac Power Board, Volume Display Board, & Tank Valve Motor - Stryker Neptune 2 ROVER ULTRA Service And Installation Manual

Waste management system, domestic/ international 120v/230v
Table of Contents

Advertisement

AC Power Board, Volume Display Board, & Tank Valve Motor
Component
AC Power Board
Volume Display Board
Tank Valve Motor
Print Date: Dec 16, 2014 04:03:57 PM
0000026959, Rev. C Effective Date: Dec 16, 2014 3:43:21 PM
Measurement
Black
Location
Lead
J2 on Backplane
Pin 12
Fuse
F3 on Board
end
Fuse
F4 on Board
end
LED D2
n/a
Pin 6
J1 on Board
Pin 6
Pin 5,
Brown
J1 - Motor Side
Pin 5,
Brown
J1 - Wire Harness
Pin 1,
Side
Black
Red
Unit of
Reading
Lead
Measure
Pin 1
120
VAC
Fuse
0
Ohm
end
Fuse
0
Ohm
end
n/a
n/a
n/a
TP3
3.3
VDC
Pin 1
5
VDC
Pin 3,
5
VDC
Yellow
1.1
VDC
Pin 4,
1.1 - 2.0
VDC
Blue
1.1 - 2.0
VDC
Pin 2,
1.0 - 12.0
VDC
Red
State
AC power board removed, Rover
power on, system idle
Fuse removed, Rover power off
Fuse removed, Rover power off
LED lights when vacuum system
button is pressed (on)
Rover power on, system idle
Rover power on, system idle
Rover power on, system idle
Rover power on, system idle
Changing tank valve position,
power on (Auto)
Magnet at varying distances, power
on (Manual)
Harness disconnected, changing
tank valve position, power on
4-11