Stryker Isolibrium PE Maintenance Manual

Stryker Isolibrium PE Maintenance Manual

Support surface
Hide thumbs Also See for Isolibrium PE:

Advertisement

Quick Links

I I s s o o l l i i b b r r i i u u m m ® ® P P E E s s u u p p p p o o r r t t s s u u r r f f a a c c e e
M M a a i i n n t t e e n n a a n n c c e e M M a a n n u u a a l l
297300000000
EN
2797
2973-009-002 Rev AB.0
2021/06

Advertisement

loading

Summary of Contents for Stryker Isolibrium PE

  • Page 1 I I s s o o l l i i b b r r i i u u m m ® ® P P E E s s u u p p p p o o r r t t s s u u r r f f a a c c e e M M a a i i n n t t e e n n a a n n c c e e M M a a n n u u a a l l 297300000000 2797...
  • Page 3 Catalogue number Serial number European medical device CE mark 2797 Authorized representative in the European Community For US Patents see www.stryker.com/patents Manufacturer Safe working load Mass of equipment Protection from liquid splash I I P P X X 4 4...
  • Page 4 Medical Equipment recognized by UL LLC with respect to electric shock, fire, and mechanical hazards in accordance with ANSI/AAMI ES60601-1:2005/(R)2012, A1:2012, C1:2009/(R)2012, A2:2010/(R)2012, and CAN/CSA-C22.2 No. 60601-1 (2008) 87VL In accordance with European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) as amended, this symbol indicates that the product should be collected separately for recycling.
  • Page 5 Main power board assembly ..........................13 Cables ..................................16 Fans and cable assembly ............................16 SPI cables (main board to foot box) assembly ......................16 Jumper from bed to Isolibrium PE mattress box cable assembly - 297100560801 ............16 Troubleshooting ..............................17 Service ..................................22 Protecting against electrostatic discharge (ESD) .....................22 Cover replacement, top ............................22...
  • Page 6 W W a a r r n n i i n n g g / / C C a a u u t t i i o o n n / / N N o o t t e e D D e e f f i i n n i i t t i i o o n n The words W W A A R R N N I I N N G G , C C A A U U T T I I O O N N , and N N O O T T E E carry special meanings and should be carefully reviewed.
  • Page 7 I I n n d d i i c c a a t t i i o o n n s s f f o o r r u u s s e e The Stryker Model 297300000000 I I s s o o l l i i b b r r i i u u m m PE support surface is intended to assist in the prevention and treatment of all pressure injury or pressure ulcers (includes all stages, unstageable and deep tissue pressure injury).
  • Page 8 Model 300900000000 P P r r o o C C u u i i t t y y bed frame Stryker reserves the right to change specifications without notice. O O p p e e r r a a t t i i o o n n...
  • Page 9 D D e e s s c c r r i i p p t t i i o o n n N N u u m m b b e e r r S S u u b b s s t t a a n n c c e e o o f f v v e e r r y y h h i i g g h h c c o o n n c c e e r r n n ( ( S S V V H H C C ) ) c c h h e e m m i i c c a a l l n n a a m m e e Cadmium Diboron-trioxide...
  • Page 10 Competent authority of the European Member State where the user and/or patient is established. To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication.
  • Page 11 S S e e r r i i a a l l n n u u m m b b e e r r The serial number (A) for your support surface is located on the patient right side of the foot box on the specification label. Unzip the support surface top cover to access the specification label.
  • Page 12 Remove product from service before you perform preventive maintenance inspection. Check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more frequently based on your level of product usage. Service only by qualified personnel.
  • Page 13 Product serial number: Completed by: Date: 2973-009-002 Rev AB.0...
  • Page 14 B B l l o o c c k k d d i i a a g g r r a a m m I I s s o o l l i i b b r r i i u u m m w w i i r r i i n n g g Foot Box 2973-009-002 Rev AB.0...
  • Page 15 P P n n e e u u m m a a t t i i c c 2973-009-002 Rev AB.0...
  • Page 16 C C i i r r c c u u i i t t b b o o a a r r d d s s P P o o w w e e r r s s u u p p p p l l y y a a s s s s e e m m b b l l y y 297100220910 Rev AF (Reference only) ISOLIBRIUM POWER SUPPLY 297100220910...
  • Page 17 M M a a i i n n p p o o w w e e r r b b o o a a r r d d a a s s s s e e m m b b l l y y 2971-022-140 Rev L (Reference only) 2973-009-002 Rev AB.0...
  • Page 18 C C a a b b l l e e a a n n d d t t e e s s t t p p o o i i n n t t V V o o l l t t a a g g e e P P o o s s i i t t i i v v e e l l e e a a d d N N e e g g a a t t i i v v e e l l e e a a d d D D e e s s c c r r i i p p t t i i o o n n...
  • Page 19 C C a a b b l l e e a a n n d d t t e e s s t t p p o o i i n n t t V V o o l l t t a a g g e e P P o o s s i i t t i i v v e e l l e e a a d d N N e e g g a a t t i i v v e e l l e e a a d d D D e e s s c c r r i i p p t t i i o o n n...
  • Page 20 C C a a b b l l e e s s F F a a n n s s a a n n d d c c a a b b l l e e a a s s s s e e m m b b l l y y 2971-021-188 Rev C (Reference only) I I t t e e m m C C o o l l o o r r...
  • Page 21 T T r r o o u u b b l l e e s s h h o o o o t t i i n n g g P P r r o o b b l l e e m m P P o o s s s s i i b b l l e e C C a a u u s s e e S S o o l l u u t t i i o o n n Support surface will not inflate...
  • Page 22 P P r r o o b b l l e e m m P P o o s s s s i i b b l l e e C C a a u u s s e e S S o o l l u u t t i i o o n n Pod manifold connector loose or Make sure that the pod manifold...
  • Page 23 P P r r o o b b l l e e m m P P o o s s s s i i b b l l e e C C a a u u s s e e S S o o l l u u t t i i o o n n Turn bladder possible leak Listen for air flows into the turn bladder...
  • Page 24 P P r r o o b b l l e e m m P P o o s s s s i i b b l l e e C C a a u u s s e e S S o o l l u u t t i i o o n n Look for a leak in the turn bladder If you find a leak, replace the turn...
  • Page 25 P P r r o o b b l l e e m m P P o o s s s s i i b b l l e e C C a a u u s s e e S S o o l l u u t t i i o o n n Check the VDC voltage at the test 1.
  • Page 26 • Do not place unprotected circuit boards on the floor. N N o o t t e e - - Always ship the circuit boards back to Stryker. Use the antistatic bag that the new board was originally shipped in.
  • Page 27 F F i i g g u u r r e e 1 1 – – T T o o p p c c o o v v e e r r 8. Insert a zipper pull tool or equivalent through the hole in the zipper slider to unlock. 9.
  • Page 28 F F i i g g u u r r e e 2 2 – – B B o o t t t t o o m m s s u u p p p p o o r r t t s s u u r r f f a a c c e e c c o o v v e e r r 9.
  • Page 29 F F i i g g u u r r e e 3 3 – – P P o o d d V V e e l l c c r r o o ® ® a a n n d d s s n n a a p p r r e e t t a a i i n n e e r r s s 17.
  • Page 30 F F i i g g u u r r e e 4 4 – – G G e e l l c c r r i i b b , , p p i i l l l l o o w w , , a a n n d d q q u u i i c c k k d d i i s s c c o o n n n n e e c c t t f f i i t t t t i i n n g g 5.
  • Page 31 F F i i g g u u r r e e 6 6 – – P P o o d d s s e e n n s s o o r r h h o o s s e e s s a a n n d d p p u u r r s s e e l l o o c c k k w w i i r r e e t t i i e e 9.
  • Page 32 N N o o t t e e • During reinstall, make sure to align the V V e e l l c c r r o o ® retainers. • After installation, remove the zipper pull tool or equivalent from the zipper. •...
  • Page 33 F F i i g g u u r r e e 8 8 – – T T o o p p c c o o v v e e r r 9. Fold the cover up toward the head end. 10.
  • Page 34 F F i i g g u u r r e e 9 9 – – G G e e l l c c r r i i b b 13. Using a T15 Torx driver, remove the nine ground screws (B) and the nine star washers (D) (Figure 10). Save the ground screws and the star washers.
  • Page 35 F F i i g g u u r r e e 1 1 0 0 – – F F o o o o t t b b o o x x c c o o v v e e r r 15.
  • Page 36 F F i i g g u u r r e e 1 1 1 1 – – S S o o l l e e n n o o i i d d v v a a l l v v e e 4.
  • Page 37 F F i i g g u u r r e e 1 1 2 2 – – P P o o w w e e r r b b o o a a r r d d N N o o t t e e - - During reinstall, torque (C) to 8 ± 2 in-lb. 4.
  • Page 38 F F i i g g u u r r e e 1 1 3 3 – – M M a a i i n n b b o o a a r r d d 4. Using needle nose pliers, apply a slight upward pressure on the main board to unclip the seven plastic standoffs. N N o o t t e e - - During reinstall, align the standoffs to the main board and press near the standoffs to seat the main board.
  • Page 39 F F i i g g u u r r e e 1 1 4 4 – – P P o o w w e e r r / / d d a a t t a a c c a a b b l l e e c c o o n n n n e e c c t t o o r r 4.
  • Page 40 5. Turn off the battery disconnect switch to turn the bed off. 6. See Accessing the foot box cover (page 28). 7. Using diagonal pliers, cut the cable tie (AY) that secures the power/data cable (BC) to the foot box. Discard the cable tie. N N o o t t e e - - During installation, tighten the cable tie (AY) to 25 - 40 lb.
  • Page 41 F F i i g g u u r r e e 1 1 6 6 – – B B o o t t t t o o m m c c o o v v e e r r N N o o t t e e •...
  • Page 42 F F i i g g u u r r e e 1 1 7 7 – – P P u u m m p p a a s s s s e e m m b b l l y y 5.
  • Page 43 L L o o w w A A i i r r L L o o s s s s ( ( L L A A L L ) ) f f a a n n r r e e p p l l a a c c e e m m e e n n t t T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
  • Page 44 F F i i g g u u r r e e 1 1 9 9 – – G G e e l l c c r r i i b b 12. Remove the two purse lock wire tie (M) (Figure 20). Save the purse lock wire ties. F F i i g g u u r r e e 2 2 0 0 –...
  • Page 45 F F i i g g u u r r e e 2 2 1 1 – – L L A A L L f f a a n n 15. Reverse steps to reinstall. N N o o t t e e •...
  • Page 46 F F i i g g u u r r e e 2 2 2 2 – – C C o o o o l l i i n n g g f f a a n n a a n n d d c c a a g g e e 4.
  • Page 47 3. Disconnect all six of the pod sensor hoses from the foot box. Push in on the coupling and pull each of the hoses out. N N o o t t e e • Do not bend or kink the pod sensor hoses. •...
  • Page 48 P P r r o o c c e e d d u u r r e e : : 1. See Accessing the foot box cover (page 28). 2. Using diagonal pliers, cut the cable ties (AY) from the inlet and outlet hoses of the resonator (AE) (Figure 24). Discard the cable ties.
  • Page 49 T T u u r r n n b b l l a a d d d d e e r r a a s s s s e e m m b b l l y y r r e e p p l l a a c c e e m m e e n n t t T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
  • Page 50 F F i i g g u u r r e e 2 2 6 6 – – P P o o d d V V e e l l c c r r o o ® ® a a n n d d s s n n a a p p r r e e t t a a i i n n e e r r s s 8.
  • Page 51 I I s s o o l l i i b b r r i i u u m m P P E E a a s s s s e e m m b b l l y y 297300210001 Rev AD (Reference only) 2973-009-002 Rev AB.0...
  • Page 52 2973-009-002 Rev AB.0...
  • Page 53 2973-009-002 Rev AB.0...
  • Page 54 2973-009-002 Rev AB.0...
  • Page 55 2973-009-002 Rev AB.0...
  • Page 56 I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 0026-672-000 Quick release pin 2971-021-005...
  • Page 57 I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 297300220008 Footbox assembly (page 54) 0058-383-000...
  • Page 58 F F o o o o t t b b o o x x a a s s s s e e m m b b l l y y 297300220008 Rev AD (Reference only) 2973-009-002 Rev AB.0...
  • Page 59 2973-009-002 Rev AB.0...
  • Page 60 2973-009-002 Rev AB.0...
  • Page 61 2973-009-002 Rev AB.0...
  • Page 62 2973-009-002 Rev AB.0...
  • Page 63 2973-009-002 Rev AB.0...
  • Page 64 2973-009-002 Rev AB.0...
  • Page 65 2973-009-002 Rev AB.0...
  • Page 66 I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 0004-880-000 Button head cap screw, Torx with star washer...
  • Page 67 I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 0038-330-000 Extension spring 0052-916-000...
  • Page 68 I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 297100560803 Cable assembly, I I s s o o l l i i b b r r i i u u m m PE power supply to Isolibrium main controller...
  • Page 69 P P u u m m p p a a s s s s e e m m b b l l y y k k i i t t - - 2 2 9 9 7 7 3 3 0 0 7 7 0 0 0 0 0 0 0 0 0 0 1 1 Rev AA (Reference only) I I t t e e m m N N u u m m b b e e r r...
  • Page 70 V V a a l l v v e e m m a a n n i i f f o o l l d d a a s s s s e e m m b b l l y y k k i i t t - - 2 2 9 9 7 7 1 1 - - 7 7 0 0 0 0 - - 0 0 0 0 7 7 Rev A (Reference only) I I t t e e m m N N u u m m b b e e r r...
  • Page 71 E E M M C C I I n n f f o o r r m m a a t t i i o o n n R R e e c c o o m m m m e e n n d d e e d d s s e e p p a a r r a a t t i i o o n n d d i i s s t t a a n n c c e e s s b b e e t t w w e e e e n n p p o o r r t t a a b b l l e e a a n n d d m m o o b b i i l l e e R R F F c c o o m m m m u u n n i i c c a a t t i i o o n n e e q q u u i i p p m m e e n n t t a a n n d d t t h h e e M M o o d d e e l l 2 2 9 9 7 7 3 3 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 I I s s o o l l i i b b r r i i u u m m P P E E s s u u p p p p o o r r t t s s u u r r f f a a c c e e The Model 297300000000 I I s s o o l l i i b b r r i i u u m m PE support surface is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
  • Page 72 The Model 297300000000 I I s s o o l l i i b b r r i i u u m m PE support surface was evaluated using the following cables: L L e e n n g g t t h h ( ( m m ) ) C C a a b b l l e e Isolibrium to bed G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ’...
  • Page 73 R R e e c c y y c c l l i i n n g g p p a a s s s s p p o o r r t t 2 2 9 9 7 7 3 3 0 0 0 0 2 2 2 2 0 0 0 0 0 0 8 8 Rev AD (Reference only) R R e e c c y y c c l l a a b b l l e e p p a a r r t t n n u u m m b b e e r r I I m m p p o o r r t t a a n n t t i i n n f f o o r r m m a a t t i i o o n n...
  • Page 74 2 2 9 9 7 7 3 3 0 0 0 0 2 2 2 2 0 0 0 0 0 0 8 8 Rev AD (Reference only) R R e e c c y y c c l l a a b b l l e e p p a a r r t t n n u u m m b b e e r r I I m m p p o o r r t t a a n n t t i i n n f f o o r r m m a a t t i i o o n n Q Q u u a a n n t t i i t t y y I I t t e e m m...
  • Page 75 2 2 9 9 7 7 3 3 0 0 0 0 2 2 2 2 0 0 0 0 0 0 8 8 Rev AD (Reference only) R R e e c c y y c c l l a a b b l l e e p p a a r r t t n n u u m m b b e e r r I I m m p p o o r r t t a a n n t t i i n n f f o o r r m m a a t t i i o o n n Q Q u u a a n n t t i i t t y y I I t t e e m m...
  • Page 76 Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 2973-009-002 Rev AB.0 2021/06 WCR: AB.3...