Stryker EOLE Operation & Maintenance Manual

Stryker EOLE Operation & Maintenance Manual

Powered support surface
Hide thumbs Also See for EOLE:

Advertisement

Quick Links

EOLE Powered Support Surface
2870
Operations/Maintenance Manual
2013/10 A.0
2870-009-001 REV. A
www.stryker.com

Advertisement

Table of Contents
loading

Summary of Contents for Stryker EOLE

  • Page 1 EOLE Powered Support Surface 2870 Operations/Maintenance Manual 2013/10 A.0 2870-009-001 REV. A www.stryker.com...
  • Page 3: Table Of Contents

    International Addresses ............. . . www.stryker.com...
  • Page 4: Symbols And Definitions

    First Digit (Solids) Protected against touch by fingers (>12.5mm); Second Digit (Liquids) IP24 Water splashing against the enclosure from any direction shall have no harmful effect. Authorized representative in the European community Catalogue Number (model) Serial Number Do Not Open with Cutter Return To Table of Contents 2870-009-001 REV. A www.stryker.com...
  • Page 5: Warning/Caution/Note Definition

    Provide special information to make maintenance easier or important instructions clearer. Return To Table of Contents www.stryker.com 2870-009-001 REV. A...
  • Page 6: Technical Specification

    • Not suitable for use in the presence of a flammable anesthetic mixture (No AP or APG protection) Air Mattress Specification Model Eole Model Number 2870 Flame Retardant Standards EN 597-1 and EN 597-2, Crib 5 Safe Working Load 160 kg / 353 lb Dimension (L x W x H) 190.5 x 89 x 18 cm / 75”x 35”...
  • Page 7: Introduction

    The safe working load for Eole is 160 kg /353 lb; the patient must not exceed safe working load specified by the support surface, frame, and accessories. Patients shall meet the minimum age requirement of 2 years old.
  • Page 8: Contact Information

    Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 Please have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication. PrODuCT SEriaL numbEr LOCaTiOn/iDEnTifiCaTiOn The serial number (A) is located at the mattress cover ▼...
  • Page 9: Summary Of Safety Precautions

    Mattress contains metal snap buttons and delrin zippers and should not be exposed under X-rays entirely. Always use X-ray cassette during portable X-ray procedure nOTE The Eole DC support surface must be used with a mattress cover at all times. The support surface cover may interact with all external skin. Return To Table of Contents www.stryker.com...
  • Page 10: Product Description

    2. Front Panel 3. Power Socket STEuErgErÄT - HinTEnanSiCHT ◄ Figure 3 4. Hanger 5. CPC Connector ◄ Figure 4 6. Eole Mattress 7. Foam and Air Cell 8. Air Hose 9. CPR Strap BEDIENFELD ◄ Figure 5 10. Lock/Unlock 11.
  • Page 11: Instructions

    Adjust the pressure control to a higher setting. Wait 10 minutes and repeat the Hand Check. If the Hand Check fails, check that the hoses are not kinked or pinched. If repeated Hand Check fails and hoses are not kinked, contact Stryker for further instruction. Warning Deflate before CPR or CPR could be ineffective.
  • Page 12: Product Functions

    Patient or caregiver can hold the lock button 3 seconds to activate or deactivate lock mode. In lock mode, Patient or caregiver can press Maxfirm/Seat button for maximum inflation. The panel of control unit will be lock automatically without any operation after 3 minutes. Return To Table of Contents 2870-009-001 REV. A www.stryker.com...
  • Page 13: Transport Mode

    4. Roll from the head end towards the foot end. 5. The power cord could be wrapped around the pump bumper on the back of pump. 6. Place the whole system into the carrying bag. Return To Table of Contents www.stryker.com 2870-009-001 REV. A...
  • Page 14: Cleaning And Disinfection

    Apply disinfectants to the external surfaces of the control unit, hoses and mattress middle layer and bottom base by wiping. Stryker recommends a chlorine-based solution with a concentration less than or equal to 1000 ppm or 70% alcohol twice a week.
  • Page 15: Trouble Shooting

    Make sure the air hose is not kinked, cracked, or split. Verify that the power Air cells fail to inflate switch is illuminated, signifying the control unit has power. Verify that the air hoses are fully inserted with a positive connection. Return To Table of Contents www.stryker.com 2870-009-001 REV. A...
  • Page 16: Service Information

    1. Disconnect the hose from control unit and mattress. 2. Discard the old hose. 3. Connect the new hose to control unit and mattress. 4. Verify proper operation of the unit before returning it to service. Return To Table of Contents 2870-009-001 REV. A www.stryker.com...
  • Page 17: Cpc Tube Replacement

    2. Verify proper operation of the unit before returning it to service. Warning Any replacement of non-authorized or wrong parts may cause the unpredictable risk rise. Please check the replaced part is suitable for Stryker Medical’s EOLE Powered Support Surface, Model 2870. Return To Table of Contents www.stryker.com...
  • Page 18: Preventative Maintenance

    Preventative Maintenance Preventative maintenance should be performed annually, at a minimum. A preventative maintenance program should be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently based on the usage level of the product.
  • Page 19 The parts and accessories listed on this page are currently available for purchase. Some of the parts identified on the assembly drawing parts in this manual may not be individually available for purchase. Please call Stryker Customer service USA at 1-800-327-0770 for availability and pricing. Product...
  • Page 20: Appendix A: Emc Information

    Power frequency magnetic fields should be at levels Power frequency characteristic of atypical (50/60Hz) magnetic 3 A/m 3 A/m location in a typical field IEC61000-4-8 commercial or hospital environment. Return To Table of Contents 2870-009-001 REV. A www.stryker.com...
  • Page 21: Guidance And Manufacturer's Declaration - Electromagnetic Immunity

    If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device. 4. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Return To Table of Contents www.stryker.com 2870-009-001 REV. A...
  • Page 22: Recommended Separation Distances

    At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. Return To Table of Contents 2870-009-001 REV. A www.stryker.com...
  • Page 23: Warranty

    Stryker, found to be effective. If requested by Stryker, products or parts for which a warranty claim is made shall be returned repaid to the factory. Any improper use or any alteration or repair by others in such manner as in Stryker’s judgment affects the product...
  • Page 24: International Addresses

    Phone: 48-22-434-88-50 Fax: 35-89-774-46820 Fax: 48-22-434-88-60 EASTERN EUROPE FRANCE PORTUGAL Stryker SA - Export Business Stryker France S.A.S. Stryker Portugal Produtos Medicos, LTDA. Grand-Rue 90 ZAC - Avenue de Satolas Green Avenida Marechal Gomes P.O. Box 1567 69330 Pusignan da Costa, 35...
  • Page 25 Stryker Medical 3800 E. Centre Avenue Portage, Michigan 49002 Stryker EMEA RA/QA Director ZAC - Avenue de Satolas Green 69881 MEYZIEU Cedex France 2013/10 2870-009-001 REV. A www.stryker.com 836001-5210 V1.0...

This manual is also suitable for:

2870

Table of Contents