Stryker Air I Operation & Maintenance Manual

Stryker Air I Operation & Maintenance Manual

Low air loss therapy mattress

Advertisement

Stryker Air I
Stryker Air II
Low Air Loss
Therapy Mattress
(LAL)
Model 2236
Operations/Maintenance Manual
Stryker Air I
(LAL Overlay)
Stryker Air II
(LAL Mattress)
For Parts or Technical Assistance:
USA: 1-800-327-0770 (option 2)
2009/05
2236-009-101 REV B
www.stryker.com

Advertisement

Table of Contents
loading

Summary of Contents for Stryker Air I

  • Page 1 Stryker Air I Stryker Air II Low Air Loss Therapy Mattress (LAL) Model 2236 Operations/Maintenance Manual Stryker Air I (LAL Overlay) Stryker Air II (LAL Mattress) For Parts or Technical Assistance: USA: 1-800-327-0770 (option 2) 2009/05 2236-009-101 REV B www.stryker.com...
  • Page 3: Table Of Contents

    Warranty................28 www.stryker.com...
  • Page 4: Symbols And Definitions

    Note This provides special information to make maintenance easier or important instructions clearer. Return To Table of Contents 2236-009-101 REV B www.stryker.com...
  • Page 5: Introduction

    INTENDED USE This manual is designed to assist you with the safe operation and maintenance of the Stryker Air I and Stryker Air II, Low Air Loss (LAL) mattress system. Carefully read this manual thoroughly before using the equipment or beginning maintenance on it.
  • Page 6: Specifications

    Ambient Temperature 100% Relative Humidity (Non-Condensing) Stryker reserves the right to change specifications without notice. Specifications listed are approximate and may vary slightly from unit to unit or by power supply fluctuations. Return To Table of Contents 2236-009-101 REV B...
  • Page 7: Therapy Mattress System Illustration

    Introduction THERAPY MATTRESS SYSTEM ILLUSTRATION Stryker Air I (LAL Overlay) Therapy Control Unit Stryker Air II (LAL Mattress) THERAPY CONTROL UNIT ILLUSTRATION Return To Table of Contents www.stryker.com 2236-009-101 REV B...
  • Page 8: Product Features

    Introduction PRODUCT FEATURES The Stryker Air I and II are comprised of two components: Therapy Control Unit Therapy Mattress System Therapy Control Unit Features • True low air loss with up to 100 liters of airflow per minute. • Three modes of operation – Autofirm, Therapy and Pulsate.
  • Page 9: Indications For Use

    The Stryker AIr I and II mattress system has safety features to prevent deflation of the therapy cells and to keep patients from bottoming out at all times. The use of the Stryker AIr I and II mattress system for these patients should be considered on an individual basis and discussed with the attending physician.
  • Page 10: Key Areas In The Treatment Of Compromised Skin

    When a patient’s skin rubs against another surface, the result is friction. The top surface of the Stryker AIr I and II mattress system therapy pad is constructed from a very smooth nylon fabric with low friction and shear properties to protect the patient’s skin from these damaging forces.
  • Page 11: Summary Of Safety Precautions

    • When using the Stryker AIr I and II mattress system, always ensure that the patient is positioned properly within the confines of the bed. Do not let any extremities protrude over the side or between the bed rails when the mattress is being used.
  • Page 12: Bed System Entrapment Information

    For a higher level of assurance, an electrician should be requested to thoroughly test the electrical system with more reliable equipment. If repair or replacement of the cord or plug is necessary, please contact Stryker at 1-800-327-0770. Figure 2...
  • Page 13: Setup Procedures

    Do not place the control unit on the floor. Position the power cord to keep personnel from tripping over it. When the product is not in use, properly store the power cord. Failure to do so could result in personal injury. Stryker Air I (LAL Mattress) Remove the existing mattress from the bed.
  • Page 14 Stryker Air I (LAL Overlay) Place the Stryker Air II LAL overlay on top of the mattress with the hose connection at the foot end of the bed and the therapy cells facing up. Secure the straps around the head and foot ends of the mattress. Fasten the two straps by the seat section securely to the sleep deck of the bed frame.
  • Page 15: Operation Guide

    Comfort Adjust The Comfort Adjust function is located on the right side of the control panel. The Stryker Air I and II can be customized to meet individual patient needs within a therapeutic window. Use the up and down keys to simultaneously increase or decrease pressure in all three zones (head, seat and foot).
  • Page 16: Cpr

    Note: Side bolsters are meant to be used with side rails and to provide a documentable and functional intervention for the risk management issues of falls and entrapment but in no way guarantees the prevention of falls or entrapment occurrences. Return To Table of Contents 2236-009-101 REV B www.stryker.com...
  • Page 17: Nursing Procedures

    RECOMMENDED LINEN Special linens are not necessary for the Stryker Air I and II. While there is no need for a bottom sheet the therapy pad should be covering the therapy cells at all times. The patient should never be lying directly on the therapy cells.
  • Page 18: Cleaning, Storage And Care

    Allow all components to air dry. Wrap mattress in plastic and return to storage area. • Thoroughly wipe down outside of control unit and allow to air dry. Cover with plastic and return to storage area • Remove gloves and dispose; wash hands. Return To Table of Contents 2236-009-101 REV B www.stryker.com...
  • Page 19: Therapy Pad

    Support Surface: Gently roll up the support surface, expelling any residual air, for temporary storage. The mattress should be wrapped in plastic and/or a clean bag for storage. Return To Table of Contents www.stryker.com 2236-009-101 REV B...
  • Page 20: Preventative Maintenance

    _____ Plastic on the control unit assembly is not damaged. _____ Mattress holds air and all straps are intact. _____ Inspect all product labeling for signs of degradation. Unit Serial Number: Completed by: _______________________________________ Date: _________________ Return To Table of Contents 2236-009-101 REV B www.stryker.com...
  • Page 21: Troubleshooting

    Ensure that the power is not on Standby. If on Standby, press the Power button. Unable to change therapy mode Make sure the Lockout function is disabled. To disable, press or adjust comfort control. the Up and Down comfort adjust arrows simultaneously. Return To Table of Contents www.stryker.com 2236-009-101 REV B...
  • Page 22: Quick Reference Replacement Parts List

    Quick Reference Replacement Parts List Note The parts and accessories listed on this page are all currently available for purchase. Please call Stryker Customer Service (800)-327-0770. Stryker Air I/Stryker Air II Power Control Unit Replacement Parts 2236-100-005 SYK Part Part Name...
  • Page 23 Quick Reference Replacement Parts List Stryker Air I/Stryker Air II Power Control Unit Replacement Parts 2236-100-005 SYK Part Part Name Part View Number Bed Hook Assembly Kit, includes 2236-100-014 screws, 2 bed hooks, springs. Power Cord Kit 2236-100-037 Mattress Hose Connector...
  • Page 24 Quick Reference Replacement Parts List Stryker Air I/Stryker Air II Power Control Unit Replacement Parts 2236-100-005 SYK Part Part Name Part View Number Blower Wire Harness Assembly 2236-100-071 Connector Hub Assembly 2236-100-062 Rubber Connector Hub Assembly 2236-100-073 Hinge Kit, includes screws 2236-100-031 and hinge set.
  • Page 25 Quick Reference Replacement Parts List Stryker Air I Mattress Replacement Parts 2236-100-003 (Reference Only) Part Name Part View SYK Part Number Hose Clamp 2236-100-027 Loop Strap 2236-100-035 Mattress Tub (Bottom Cover) 2236-100-033 Therapy Pad (Top Cover) 2236-100-034 Foot section hose assembly...
  • Page 26 Quick Reference Replacement Parts List Stryker Air II Mattress Replacement Parts 2236-100-004 (Reference Only) Part Name Part View SYK Part Number Air Cell, LAL (6”x36”) 2236-100-044 Air Cell, Standard CairRail 2236-100-045 Back Raise Sensor 2236-100-026 Assembly (with Plate) External Hose Assembly...
  • Page 27 Quick Reference Replacement Parts List Stryker Air II Mattress Replacement Parts 2236-100-004 (Reference Only) Part Name Part View SYK Part Number Therapy Pad (Top Cover) 2236-100-043 Foot section hose assembly Seat section hose assembly Head section hose assembly Head Section:...
  • Page 28: Warranty

    LIMITED WARRANTY The Stryker Air I & II mattress system is guaranteed for a period of one year from the date of delivery, against defects in materials and workmanship, under normal use and service. This one-year warranty includes all mechanical and electrical components.
  • Page 30 UNITED STATES UNITED STATES Stryker Medical Stryker Medical 3800 E. Centre Ave., 3800 E. Centre Ave., Portage, Michigan USA Portage, Michigan USA 49002 49002 CANADA CANADA Stryker Canada Stryker Canada 45 Innovation Drive 45 Innovation Drive Hamilton, Ontario Canada Hamilton, Ontario Canada...

This manual is also suitable for:

Air iiLal overlayLal mattress2236-000-0022236-000-003

Table of Contents