Electromagnetic Compatibility (Emc) - Miele PLW 8615 Safety Operating Instructions Manual

Washer-disinfector
Hide thumbs Also See for PLW 8615 Safety:
Table of Contents

Advertisement

Electrical connection
84

Electromagnetic compatibility (EMC)

The machine has been tested for electromagnetic compatibility (EMC)
in accordance with EN 61326-1 and is suitable for operation in com-
mercial environments, such as hospitals, medical practices and labor-
atories and other similar environments which are connected to the
mains power supply.
The machine's high frequency (HF) energy emissions are very low and
are therefore unlikely to interfere with other electronic machines in the
vicinity.
Flooring in the installation area must be wood, concrete or tiled. Syn-
thetic flooring must be able withstand a relative humidity level of
30 % to minimise the risk of electrostatic discharges.
The quality of the power supply should comply with that found in a
typical commercial or hospital environment. Check that the power
supply voltage is within a range of +/-10 % (UK: -6/+10 %) of its
nominal value.