Miele PLW 8617 Operating Instructions Manual
Miele PLW 8617 Operating Instructions Manual

Miele PLW 8617 Operating Instructions Manual

Laboratory glassware washer
Hide thumbs Also See for PLW 8617:
Table of Contents

Advertisement

Quick Links

Operating Instructions
Laboratory Glassware Washer
PLW 8617
To prevent accidents and machine damage,
en - US, CA
read these instructions before installation or
use.
M.-Nr. 12 176 860

Advertisement

Table of Contents
loading

Summary of Contents for Miele PLW 8617

  • Page 1 Operating Instructions Laboratory Glassware Washer PLW 8617 To prevent accidents and machine damage, en - US, CA read these instructions before installation or use. M.-Nr. 12 176 860...
  • Page 3: Table Of Contents

    Contents Guide to the manual ......................Appropriate use ....................... User profiles........................Machine description......................Machine overview......................Control elements on the display..................11 Main menu........................11 Program start display ....................13 IMPORTANT SAFETY INSTRUCTIONS ................18 Symbols on the machine....................23 Main menu........................24 Main menu.........................
  • Page 4 Contents Program sequence ......................51 End of program........................52 Canceling a program ......................52 Program cancellation with manual disinfection............. 52 Delay start ......................... 53 Network connection......................54 Network connection ......................54 Network protocols ......................55 Process documentation ....................55 System settings ....................... 56 System ..........................
  • Page 5 Contents After sales service......................88 Electrical connection ...................... 89 Electromagnetic compatibility (EMC) ................90 Plumbing .......................... 91 Technical details ......................92 Caring for the environment .................... 94 Appendix .......................... 96 User rights in the Main menu .................... 96 User rights in the System menu ..................97 Power levels of the programs/Mobile units/Module combinations ........
  • Page 6: Guide To The Manual

    Guide to the manual Warnings  Information which is important for safety is highlighted in a thick framed box with a warning symbol. This alerts you of potential dan- ger of injury to people or damage to property. Read these warning notes carefully and observe the instructions and codes of practice described.
  • Page 7: Appropriate Use

    Observe the information issued by the manufacturer of the wash items. Inappropriate use The PLW 8617 Laboratory Glassware Washer may not be used for any purposes other than the appropriate use described. This applies in particular to: - the reprocessing of medical devices suitable for reprocessing...
  • Page 8: User Profiles

    User profiles User administration Access to the machine is restricted to a certain group of people. Au- thorized users are entered in User management via the machine’s web interface. User management is described in the programming manual. A User ID and PIN are required for the user Login on the machine. User actions such as starting a reprocessing program are saved.
  • Page 9: Machine Description

    Machine description Machine overview Infeed and out- feed side Service cover for Miele Technical Service I/0 power switch Display Docking system for transfer trolleys, service cover lock Service cover Door closed...
  • Page 10 Infeed and out- feed side without service cover Open door Water supply for load carriers Filter combination USB port Connection for Miele Technical Service Drawer with containers and connections for process chemicals Data plate Lower machine spray arm Upper machine spray arm...
  • Page 11: Control Elements On The Display

    Machine description Control elements on the display Main menu The machine display shows the control panel light on a black back- ground. In these operating instructions, the display is shown as black on a light background in order to improve legibility. ...
  • Page 12 Machine description Symbols in the Symbol Description/function of the button main menu Opens the list of reprocessing programs  Reprocessing which are assigned to the load carrier  Favorites Opens the list of favorites  Documentation Opens the list of available wash reports Opens the configuration menu, e.g., for ...
  • Page 13: Program Start Display

    Machine description Program start dis- Once the A 500 load carrier, for example, has been selected in the play Reprocessing menu and the associated Universal reprocessing pro- gram has been confirmed, the following is displayed:  user/  Universal 18.06.2018 15:05  1 : 2 6 Time left ...
  • Page 14  Door locked  Shows the program sequence display in the default view Schematic program sequence display, contains informa- In/Out tion for Miele Technical Service  Displays the program sequence as a diagram  Selects the steam heating method ...
  • Page 15 Machine description °C Example of a pro- After the program has started, the following display is shown: gram sequence    Universal user/ display 18.06.2018 15:05 Program block Main wash 1  0 : 2 5 167 °F Temperature Duration The current program phase and the temperature reached are also shown on the display.
  • Page 16 Machine description Once you have switched to the table view using the  button, the Program cycle dis- play – table view following data is displayed: - Block name - Program phase countdown - Holding time - Program parameters with the associated target and actual values and the minimum and maximum values reached It is possible to scroll to other pages using the scroll bar.
  • Page 17 - Air temperature - Wash pressure at the circulation pump - Conductivity This information is used by Miele Technical Service and is not ex- plained in these operating instructions. Graphical program Once you have switched to the graphical view of the program se- sequence display quence using the  button, the following values are displayed as...
  • Page 18: Important Safety Instructions

    (e.g., temperature, moisture, chemical resistance, abra- sion resistance, and vibration). This special bulb must only be used for the purpose for which it is intended. It is not suitable for room lighting. Replacement bulbs may only be installed by Miele Technical Service. Risk of injury Please pay attention to the following notes to avoid injury.
  • Page 19 Label machines which have been taken out of operation and lock them to prevent them being switched on again without authorization. The machine may only be put back into operation once it has been successfully repaired by Miele Technical Service.  Personnel operating the machine should be trained regularly. Un- trained personnel must not be allowed access to the machine or its controls.
  • Page 20 Process chemicals can, in certain circumstances, cause damage to the machine. Always follow the recommendations of the process chemical manufacturer. In case of damage or doubt about compatibility, please contact Miele.  Cleaning detergents containing chlorine can damage the elas- tomers of the machine.
  • Page 21 In critical applications where very stringent requirements have to be met, it is strongly recommended that all the relevant factors for the process, such as process chemicals, water quality, etc., are dis- cussed with Miele.  If the cleaning result is subject to particularly stringent require- ments, e.g., in chemical analysis, regular quality control should be...
  • Page 22 IMPORTANT SAFETY INSTRUCTIONS  The amount of residual solvents and acids on items going into the cabinet should be minimal. There should be no more than a trace of any solvents with a flash point of below 70°F (21°C).  Chloride solutions, in particular hydrochloric acid, must not be placed in the cabinet.
  • Page 23: Symbols On The Machine

    Using components and accessories  Only Miele accessories should be connected to this machine. They must be suitable for the application they are used for. Consult Miele for details on the type of accessories that can be used.  Only use Miele Mobile Units, Modules, and Inserts. Using mobile...
  • Page 24: Main Menu

    Main menu Main menu    Login Main menu 18.06.2018 15:05   Favorites Reprocessing   Documentation System  Actions in The main menu of the machine is split into headers and footers and the Main menu 4 menus. The following actions can be performed using the buttons: - Lock the display, e.g., for surface cleaning (see “Maintenance, lock- ing machines”) - Change the display language...
  • Page 25: Changing The Display Language

    Main menu Changing the dis- Changing the display language has no impact on the set system lan- play language  guage. Reports and printouts are issued in the system language.   Press the  button in the header of the display. ...
  • Page 26: Logout

    Main menu Logout The logged-in user is automatically logged out after 24 hours if no button has been pressed on the display in that time, or if the machine carries out an automatic restart. The automatic restart can be set in the web interface under “System, Automatic restart”.
  • Page 27 Main menu Download All available reprocessing reports can be downloaded to a USB stick. The USB port is located behind the service cover on the infeed side of the machine. The USB stick must be FAT-32-formatted and have a minimum ca- pacity of 1 GB.
  • Page 28: Opening And Closing The Door

      Main menu Opening and clos- The status of the machine door is shown in the footer of the display. ing the door  -  The door has been released and can be opened. -  The door is open and can be closed. -  The door is locked.
  • Page 29: Application Technology

    Miele offers suitable load carriers for all areas of application defined in “Appropriate use”, such as mobile units, modules, inserts, and spe- cial rinsing fittings. More information can be obtained from Miele.
  • Page 30: Arranging The Wash Items

    Application technology Arranging Only items which have been declared by their manufacturer as suit- the wash items able for machine reprocessing may be processed. The manufac- turer’s specific reprocessing instructions must be observed.  Contaminated items pose a health risk. Contaminated items can result in various hazards to health, which can lead to infections, poisoning, injuries, and more depending on the type of contamination.
  • Page 31: Preparing The Wash Items

    Application technology - Broken glass can result in serious injury when loading or unloading. Damaged glass wash items must not be reprocessed in the ma- chine. - Nickel and chrome-plated wash items and wash items made of alu- minum are not generally suitable for machine reprocessing. Special process conditions are required for these wash items.
  • Page 32 Application technology Empty all wash items before loading into the machine, paying par-  ticular attention to relevant regulations. Remove non-water soluble residues such as paint, adhesives, and  polymer compounds using appropriate solvents. Rinse wash items which have been in contact with solvents, chlo- ...
  • Page 33: Before Starting A Program

    Application technology Before starting Carry out a visual check before starting every program: a program - Is everything correctly loaded/connected for cleaning? - Can the wash water consistently permeate wash items with 2 open- ings, such as pipettes and viscometers? - Are the spray arms clean and do they rotate freely? - Is the filter combination clean? Remove any coarse soiling and clean the filter combination if nec- essary.
  • Page 34: Laboratory Glassware And Utensils

    Application technology Laboratory glassware and utensils ...wide-neck Wash items with wide necks, such as glass beakers, wide-neck Er- lenmeyer flasks, and petri dishes, or cylindrical items such as test tubes, can be cleaned and rinsed inside and out using rotating spray arms.
  • Page 35: Program Overview

    Application technology Program overview On delivery, the reprocessing programs are assigned to the A 503 mobile unit. The follow- ing modules can be used in various combinations: Exceptions to this are the Vials Pipettes  programs (see table). In order to use other mobile units, new load carriers must be set up in the web interface under “Load carrier management”.
  • Page 36: General Program Information

    Application technology General program information Program Area of application Mobile unit Mini - Very short program for barely soiled wash items and very low A 503 rinse requirements - For a wide range of soiling - Not suitable for denatured residues such as protein - Not suitable for inorganic, acid-soluble residues such as metal- lic salts - In 2 power levels...
  • Page 37: Programs For Specific Soiling

    Application technology Programs for specific soiling Program Area of application Mobile unit Inorganic - Program for lightly soiled to moderately soiled wash items and A 503 moderate to high rinse requirements - For removing inorganic, acid-soluble residues such as metallic salts - In 2 power levels Organic...
  • Page 38: Programs For Specific Wash Items

    Application technology Programs for specific wash items Program Area of application Mobile unit Plastics - Program for lightly soiled to moderately soiled plastics and A 503 moderate rinse requirements - For heat-sensitive laboratory equipment, e.g., plastic bottles that are resistant to temperatures of up to 130°F (55°C) - In 2 power levels Vials - Program for lightly soiled to moderately soiled small wash...
  • Page 39: Additional Programs

    Application technology Additional programs Program Area of application Mobile unit Laboratory - Program for cleaning and thermal disinfection at 199°F (93°C) A 503 Special and 10 min holding time in the first program phase; wash wa- ter is only pumped out after the disinfection process - In 2 power levels Rinse cold wa- - Program for rinsing wash items with cold water, e.g., to re-...
  • Page 40: Chemical Processes And Technology

    This section is intended as a guide. If unforeseen interactions occur during reprocessing, or if you have any questions on this subject, please contact Miele. General information Problem...
  • Page 41 - Check the throughput of the dispensing system regularly. - Ensure that the regular cycle of mainte- nance is observed. - Contact Miele for advice. Process chemicals can damage elastomers - Follow the process chemical manufac- and plastics in the machine and acces- turer’s instructions and recommendations.
  • Page 42 - damage to elastomers and plastics in the without de-foaming agent using the machine ganic  program - damage to certain plastics, e.g., polycar- - Contact Miele for advice. bonate and plexiglass, in the wash items being processed...
  • Page 43 Chemical processes and technology Soiling Problem Solution The following substances can damage the - Depending on usage, wipe the lower door elastomers in hoses and seals, as well as seal on the machine periodically with a the plastics in the machine: lint-free cloth or sponge.
  • Page 44 Chemical processes and technology Reaction between process chemicals and soiling Problem Solution Natural oils and fats can be emulsified with - Use the  program. alkaline process chemicals. This can lead to - Use a special program that dispenses pH- a heavy build-up of foam. neutral emulsifiers containing surfactants in the pre-rinse.
  • Page 45: Dispensing Liquid Process Chemicals

    Dispensing liquid process chemicals  Unsuitable process chemicals pose a health risk. Using unsuitable process chemicals will generally cause an unsat- isfactory wash result and can pose a health risk or cause damage to property. Only use process chemicals designed specifically for use in labora- tory glassware washers and follow the manufacturer’s instructions on how to use them.
  • Page 46: Containers For Process Chemicals

    4 external dispensing systems can be connected instead of the 4 op- tional internal dispensing systems. If a dispensing system is to dispense different process chemicals, the chemicals must be changed by Miele Technical Service or an authorized technician. Containers for The drawer for containers with process chemicals is located behind process chemi- the service cover.
  • Page 47: Filling The Dispensing System

    Dispensing liquid process chemicals Push the suction wand into the neck of the container and screw it  into place. Place the container in the drawer.  The suction wand must be located on the lower side of the support. Push the drawer into the machine.
  • Page 48: Operation

    Operation Switching the machine on and off The I/0 power switch disconnects the machine from the power sup- ply. The switch is located above the display. Press the I power switch to switch the machine on.  Once the boot process has finished, the machine is ready for opera- tion.
  • Page 49: Selecting A Program

    Operation Selecting a program The  reprocessing programs are assigned to certain load carriers. Therefore, a load carrier must be selected first with the help of a pic- togram. Press the   Reprocessing Selecting a load  button.  carrier The menu for selecting the load carrier via a pictogram is displayed. user/ Select load carrier ...
  • Page 50: Selecting Programs From Favorites

    Operation If  reprocessing programs have been defined in the   System, Man- Selecting pro- grams from age favorites  menu, the program can be selected using the   vorites Favorites  button. Press the   Favorites  button.  Select load carrier  menu is displayed. Press the button for the required load carrier, e.g.,  A 503.
  • Page 51: Starting A Program

    Operation Starting a program Check whether the right reprocessing program for the load carrier is displayed in the header. With the  button, you can open the menu for selecting the repro- cessing programs if you wish to choose a different reprocessing program.
  • Page 52: End Of Program

    Operation End of program After the program has ended, the following is displayed: - Cycle number - time elapsed since the end of the program - Result of the wash process Program finished  message The progress bar changes from white to green. The wash cabinet lighting can be switched on automatically depend- ing on the pre-setting.
  • Page 53: Delay Start

    Operation Manual disinfection Press OK to acknowledge the fault message in the list.  Press OK to acknowledge the program cancellation.  Please unload The message appears in the program sequence dis- play. Open the door on the infeed side. ...
  • Page 54: Network Connection

    (e.g., using port forwarding)! The IP addresses from 192.168.10.1 to 192.168.10.255 are re- served for Miele Technical Service. They must never be set in the internal network. The machine should be operated only in a separate network segment which meets one of the following requirements: –...
  • Page 55: Network Protocols

    Network connection Network protocols The machine supports communication via the following network pro- tocols: Network Description Port protocols DHCPv4 Assignment of dynamic IPv4 addresses 68/UDP DHCPv6 Assignment of dynamic IPv6 addresses 546/UDP HTTPS Web interface for configuring the ma- 443/TCP chine parameters The ICMPv4 and ICMPv6 protocols are also supported.
  • Page 56: System Settings

    System settings System The   System  menu contains the following functions: user/ System Network connections Machine functions Restart Data plate Manage favorites Press the buttons to open the sub-menus. The functions in the sub- menus are described on the following pages. Network connections Network connections  menu contains network configuration and...
  • Page 57: Changing The Pin

    The new PIN is applied in the User management in the web interface. It is not visible there. If a user has forgotten their PIN, they can reset it in the User management. Administrator The PIN for the  user level can only be reset by Miele Technical Service.
  • Page 58: Pairing

    System settings Pairing If the machine is connected to a network, external applications can access the machine controls, e.g., software for Process documenta- tion or the Web interface. When the machine is connected for the first time, Pairing must be carried out between the machine and the exter- nal application.
  • Page 59: Granting Write Permissions

    System settings Granting write If data needs to be downloaded via the web interface or machine set- permissions tings need to be changed, the logged-in user requires write permis- sions. As soon as a user has requested write permissions, the follow- ing message appears in the display: ...
  • Page 60: Configuring The Network

    System settings Network Configuring the The Ethernet connection is configured in the  menu. The net- network work can be configured on the machine’s display or using the web in- terface. Before the machine is connected for the first time, e.g., to a network printer, the network must be configured on the machine dis- play.
  • Page 61: Configuring The Printer

    System settings Configuring the A connected printer can be set up using the display or the web inter- printer face. To set the printer up in the web interface, see the programming manual. Printer Press the  button.  Type Press the button next to the  parameter.
  • Page 62: Certificate Presentation

    System settings Certificate pre- The machine has a self-signed TLS certificate in the x.509 standard. sentation The certificate is used whenever you access the machine via the net- work with a web browser. It is also used to authenticate secure com- munication with the machine, e.g., in conjunction with software for process documentation.
  • Page 63: Machine Functions

    System settings Machine functions Machine functions  menu contains service functions, e.g., for dis- Machine func- pensing systems and for built-in tanks. Once the tions  menu has been opened, the following menu will appear on the display:  S P / W O R L D L o g I n ...
  • Page 64: Carrying Out Self-Disinfection (Optional)

    Self-disinfection completed Self-disinfection failed Press OK to confirm the message.  If self-disinfection is not repeated successfully, contact Miele Tech- nical Service. You can cancel self-disinfection with the  button in the footer. Cancel self-disinfection?  confirmation prompt is displayed. Press...
  • Page 65: Starting The Continuous Program

    System settings Starting the con- A repeated program start can be defined for a reprocessing program tinuous program Continuous program in the  menu. This function can be used for wash- related trials, e.g., material tests, or for periodic decontamination in the event of long downtimes.
  • Page 66: Filling The Dispensing System

    If the machine is equipped with a drying unit, the coarse and fine fil- nance ters in the drying unit must be replaced by Miele Technical Service on a regular basis. You can read off the operating times and the times left for the individ-...
  • Page 67 System settings Carrying out setup Setup for a dispensing system can only be performed when there are no load carriers in the wash cabinet. To determine the precise throughput of the dispenser pump and the pulse rate of the flow sensor, 2 measuring cylinders are required: - 1 x 1,000 ml - 1 x 100 ml Put the dispensing system suction wand in a 500–1,000 ml measur-...
  • Page 68 Set up dis- OK; pressing  retains the old conversion factors. The pensing system  menu reappears in the display. Set up dispensing sys- Contact Miele Technical Service to carry out if the machine dispensing systems are not equipped with suc- tion wands.
  • Page 69: Calibrating The Conductivity Meter

    System settings Calibrating the When particularly high demands are placed on the cleaning and final conductivity me- rinse result, e.g., in chemical analysis, the conductivity meter can be Adjust conductivity module calibrated with the  function. The measuring system is calibrated from DI water to cold water in the conductivity measuring range.
  • Page 70: Carrying Out A Program Test

    System settings Carrying out a Newly created or modified reprocessing programs can be checked program test Program test with the . A program can only be enabled with the Checked  parameter if a  reprocessing program has run through the Program test without any faults (see programming manual, “Program administration”).
  • Page 71 System settings Press the button for the required reprocessing program.  The following message appears on the display: Program test: Skip program block with . Door can be opened at end of each program block. Press OK to confirm the message. ...
  • Page 72: Restarting

    System settings Restarting Restart  function shuts down the machine controls and then re- Restart boots them. This can be used, for example, when you get the Switch machine off and on prompt when you are troubleshooting. Restart Press the  button. ...
  • Page 73: Managing Favorites

    System settings Managing favorites You can define your 4 most commonly used programs as favorites in Manage favorites  menu. Automatic program allocation  is not activated: After the  Reprocessing  button has been pressed, all  reprocessing programs assigned to the load carrier are displayed. If you select the reprocessing program via the  ...
  • Page 74: Maintenance Measures

    Maintenance measures Maintenance The machine should be serviced every 2,000 hours of operation, or at least once a year, by Miele Technical Service. Maintenance covers the following points and functional checks: - electrical safety check compliant with national rules and regulations (e.g., VDE 0701, VDE 0702)
  • Page 75: Routine Checks

    Maintenance measures Routine checks Before the start of each working day, the operator must conduct a series of routine checks. A routine checklist is supplied with the ma- chine. The following items must be checked: - all filters in the wash cabinet - spray arms in the machine and in any mobile units, modules, or baskets - the wash cabinet and the door seal...
  • Page 76: Cleaning The Filters In The Wash Cabinet

    Maintenance measures Cleaning the filters in the wash cabinet The filters in the floor of the wash cabinet prevent coarse soiling from coming into contact with the circulation system. Filters can become blocked by soiling, so they need to be checked every day and cleaned as necessary.
  • Page 77: Cleaning The Spray Arms

    Maintenance measures Remove the microfine screen filter.  Remove deposits and soiling with running water, using a brush if  necessary. Blast the microfine screen filter with compressed air against the di-  rection of flow of the wash water. Insert the microfine screen filter back into the sump.
  • Page 78 If there is considerable visible wear on the spray arm bearings, con- tact Miele Technical Service as this can result in functional prob- lems. Reinsert the spray arms after cleaning, then secure them with the ...
  • Page 79: Cleaning The Machine

    Maintenance measures Cleaning the machine Locking the dis- The machine display can be locked for cleaning and surface disinfec- play tion. Press the  button in the header.  The following will appear on the display:   To unlock the machine display, hold the  button for around 3 s. ...
  • Page 80 Cleaning the The wash cabinet is mostly self-cleaning. wash cabinet However, if you notice a build-up of deposits, please contact Miele  Technical Service for advice. Cleaning the door The door seal should be cleaned regularly with a damp cloth to re- ...
  • Page 81: Checking Mobile Units, Modules, And Inserts

    Maintenance measures Checking mobile units, modules, and inserts Mobile units, modules, and inserts should be checked daily to make sure they are functioning correctly. A checklist is supplied with the machine. The following points must be checked: - Are the guide rails in good condition, and are they securely at- tached to their mobile units? - Are the lock nuts for the guide rails tightened securely? - Are the locking caps in the module connectors of the modular sys-...
  • Page 82: Built-In Printer (Optional)

    Maintenance measures Built-in printer (optional) Lever for opening the flap Paper feed Acknowledgement button LED status Printer status Power supply fault Ready for operation 1 x flash Not ready for operation 2 x flash Fault on the printer head 3 x flash Paper jam/paper roll empty 4 x flash Temperature fault Replacing the pa-...
  • Page 83: Thermal Paper

    Maintenance measures Thermal paper Replacement rolls of thermal paper with a width of 2 3/8" (60 mm) and an external diameter of approximately 2" (50 mm) can be ob- tained from the machine manufacturer. When this is stored under the following conditions, the text will re- main legible for at least 7 years: - no direct light - constant room temperature between 64°F (18°C) and 77°F (25°C)
  • Page 84: Fault Messages And Instructions

    The fault messages on the machine display are designed to help you find the cause of the malfunction and to resolve it. However, please note the following:  Repairs should only be performed by Miele Technical Service or an authorized technician. Unauthorized repairs can expose the user to considerable risk.
  • Page 85: Technical Faults On The Machine

    Fault messages and instructions Technical faults on the machine List with active   user/ Messages 18.06.2018 15:05 messages  2650 Timeout door gaskets inflation 2 8 . 0 1 . 2 0 1 9 1 1 : 0 9 ...
  • Page 86: Troubleshooting

    Fault messages and instructions Troubleshooting    user/ 2841 Fault 18.06.2018 15:05 Description Confirm message with OK. Check user administration. If fault persists, contact Service. P L W 8 6 1 7 Machine type Fabrication number 00 / 123456789 ...
  • Page 87: Technical Faults Without Fault Number

    – The display is faulty. – The connecting cable between the display and the controls is faulty. – The plug connection is faulty. Contact Miele Technical Service.  The machine is not switched on. Press the I power switch. ...
  • Page 88: After Sales Service

    Fault messages appear on the machine display. Notify Miele Technical Service if you are still not able to resolve the problem after following the advice on the machine display. Miele Technical Service will need to know the model and serial num- ber of the machine.
  • Page 89: Electrical Connection

    *INSTALLATION* Electrical connection  All electrical work must be carried out by a qualified electrician in accordance with local and national safety regulations. - The electrical installation must be in compliance with current local and national safety regulations (DIN VDE 0100 in Germany). - The connection to the power supply must comply with national reg- ulations.
  • Page 90: Electromagnetic Compatibility (Emc)

    *INSTALLATION* Electrical connection Electromagnetic compatibility (EMC) The machine has been tested for electromagnetic compatibility (EMC) in accordance with EN 61326-1 and is suitable for operation in com- mercial environments, such as hospitals, medical practices, and labo- ratories plus other similar environments which are connected to the domestic electrical supply.
  • Page 91: Plumbing

    *INSTALLATION* Plumbing  Health risk and risk of damage due to contaminated incoming water. Machine must be connected to a potable tap water supply. Please refer to the installation diagram supplied with the machine. - If the water supply has a high iron content, there is a risk of corro- sion occurring on wash items and the machine itself.
  • Page 92: Technical Details

    Technical details Height 98 1/4" 2495 mm Width 45 1/4" 1150 mm Depth 39" 990 mm Wash cabinet dimensions:   Height 26 9/16" 675 mm   Width 35 7/16" 900 mm   Depth 39" 990 mm Net weight incl. stand and top-box paneling 1052 lb 477 kg Net weight plus EL/S, DU, SC, CT * 1470 lb 667 kg Operating weight...
  • Page 93  mark Machinery Directive 2006/42/EC Manufacturer’s address Miele & Cie. KG, Carl-Miele-Str. 29, 33332 Gütersloh, Germany * If installed above 4,921 ft (1,500 m), the boiling point of the wash water will be lower. Therefore, the disinfection temper- ature and the holding time might need to be adjusted.
  • Page 94: Caring For The Environment

    Instead, please make use of officially designated collection and dis- posal points to dispose of and recycle electrical and electronic appli- ances in your local community, with your dealer, with Miele, or with the manufacturer. By law, you are solely responsible for deleting any personal data from the appliance prior to disposal.
  • Page 95 Appendix...
  • Page 96: Appendix

    Appendix User rights in the Main menu Action      Open and close the loading door      Open and close the unloading door after the end of a program depending on the Authentification - End of program setting in the web interface Select the electric or steam heating method (see “Machine ...
  • Page 97: User Rights In The System Menu

    Appendix User rights in the System menu Main menu/System     Network connections Change PIN   Pairing Network     Printer   Certificate presentation    Machine functions Self-disinfection Continuous program    ...
  • Page 98: Power Levels Of The Programs/Mobile Units/Module Combinations

    Appendix Power levels of the programs/Mobile units/Module combinations Mobile unit Modules Programs/Power levels A 500 - A 603 Mini Standard Universal Intensive Organic Inorganic Agar Plastics Laboratory Special A 501 Mini Plus Standard Plus Universal Plus Intensive Plus Organic Plus Inorganic Plus Oil Plus Agar Plus Plastics Plus Laboratory Special Plus...
  • Page 99 Appendix Mobile unit Modules Programs/Power levels A 503 1–3 modules: Mini - A 606 Standard Universal - A 612 Intensive - A 620 Organic - A 621 Inorganic - A 622 - A 623 Agar Plastics Laboratory Special A 503 4–6 modules: Mini Plus - A 606 Standard Plus Universal Plus - A 612 Intensive Plus - A 620...
  • Page 100 Oil Plus Agar Plus Plastics Plus Laboratory Special Plus E 941 - 1–4 E 947/2 Vials Plus If the E 942/3 modules are to be used for pipettes, please contact Miele. Service programs Mobile unit Modules Power levels/ program Zero - Without modules Tank emptying...
  • Page 101: Power Levels Of The Programs/Mobile Units/Module Combinations For The Aplw 511

    Appendix Power levels of the programs/Mobile units/Module combinations for the APLW 511 Cus- Programs with “Max” power level must be entered manually in the web interface as  tomer program Program control in the  . To do this, copy the corresponding program from the Plus power level and adjust the program parameters.
  • Page 102: Optional Extras

    Appendix Optional extras Equipment Standard Option Drain valve (DV) Electric heating (EL) Conductivity monitoring (CS) Spray arm monitoring (RS) Option to switch between steam/electric heating for wash cabinet and DI tank (if present) (EL/S) Recycling tank (RT) DI tank (TH) Additional options: 1–4 additional internal dispensing systems (DS) Drain pump (DP)
  • Page 104 Technical Service & Support Phone: 800-991-9380 Miele Professional proservice@mieleusa.com Technical Service Phone: 1-888-325-3957 serviceprofessional@miele.ca Manufacturer: Miele & Cie. KG Carl-Miele-Straße 29, 33332 Gütersloh, Germany Alteration rights reserved / Publication date: 2022-08-16 M.-Nr. 12 176 860 / 00 Valid for software version: 1.3...

Table of Contents