ELECTROMAGNETIC COMPATIBILITY (EMC)
Guidance and manufacturer's declaration - electromagnetic emissions
The
device is intended for use in
the
electromagnetic environments
listed below,
and should
only be used
in such environments:
Emissions teat
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
RF
energy is used only to maintain
device's
operation. Therefore.
its
RF
emissions
are
so low that it's not likely to cause
any
interference in
nearby
RF emissions CISPR 11
Class B
electronic equipment.
Harmonic emissions IEC 61000-3-2
Not
Applicable
The device is suitable
for
use in all establishments, including domestic
establishments, and those directly connected to the public low-voltage
Voltage fluctuations/ flicker
Not Applicable
power supply network that supplies buildings used for domestic
emissions IEC 61000-3-3
purposes.
Guidance and manufacturer's declaration - electromagnetic Immunity
The
device
is intended for use in the electromagnetic environments
listed below,
and should
only be used
in such environments:
Immunity teat
IEC 60601 teat level
Compliance level
Electromagnetic environment-guidance
Electrostatic
±
8 kV contact
±
8 kV contact
In the case of
air
discharge testing, the climatic
dischar~e
(ESD)
dischaoie
discharge
conditions shall be
within
the
following
ranges:
IEC 610 0-4·2
±
15 k air discharge
±
15 kV air discharge
Ambient
Temperature:15°C-35°C
Relative
Humidity:
30%~60%.
Power frequency
30A/m
30A/m
Power frequency magnetic fields should be at
magnetic field
SO or60 Hz
SO or 60 Hz
levels
characteristic of a
typical
location
In
a
IEC 61000-4-8
typical commercial or hospital environment.