JOB CARD PIST044–001 – Innova 2000 System Interconnect
Time: 9 h 00 min – Personnel: 2 field engineers = 18 h 00 of workload
1
SUPPLIES
D None.
2
TOOLS
D Standard Tools: screwdrivers, drilling machine.
3
SAFETY PRECAUTIONS
D Follow the procedure to unpack the spool.
D Attach the power strip on wall to respect regulatory constraints.
4
PREREQUISITES
D Gullets, raceways in place.
D Different subsystems shall be color coded to check the cable distribution in real time:
D Yellow for POSTIT cabinet.
D Orange for ACAB cabinet.
D Red for MPPU cabinet.
D Purple for X–Ray tube chiller & Detector conditioner – LC gantry.
D Blue for Gantry.
D Brown for Table.
Innova 2000 Cardiovascular Imaging System
143
Pre–Installation Manual
pim 2337741–100
1 of 10