Illustration 12 – THROUGH–BOLT SUPPLIED (PAN TYPE FLOOR CONSTRUCTION)
FOR EXAMPLE
For pan type floor
construction, joists must
be spanned with steel
channels (customer
furnished)
Illustration 13 – LOCATION OF FIXING BOLTS
INNOVA LC
POSITIONER
12 bolts with
a pull out of
3910 N
For alternative table bolts or seismic area, refer to template drawing shown in Illustration 43.
Hole dimension and preferred location in concrete floor
If in the examination room, the Innova LC Positioner and the OMEGA patient tables
are not placed on a computer floor but directly put on concrete floor, the location of
the cable access needs to be carefully planned.
Otherwise, if the cable run is located under the concrete floor, the cables will have to
come through the floor and in this case you will need two holes, one for the LC
Positioner and the other for the patient table.
The diameter of both holes can be the same (225 mm / 9 in).
1/4 inch (6 mm) to
1/2 inch (13 mm)
grout pad
Floor
É É É É É É É É É É É É É É É É
È È
È È
È È
È È
È È
È È
549.4  2.54
(21.63  0.10) DIA
Isocenter
LONG
(optional)
1125.7  1.5
(44.32  0.06)
TO BOLT HOLES
Innova 2000 Cardiovascular Imaging System
Baseplate
Furnished thru bolts will accommodate
8 inch (203 mm) to 17 inch (432 mm)
thick floor
Spacers
(Customer furnished)
COMPACT
& LONG
1242.6  1.5
(48.92  0.06)
45
Pre–Installation Manual
pim 2337741–100
thru bolt M20–500–400
È
È
È
È
È
È
2-hex nuts M20 STL
4 bolts with
a pull out of
5650 N
OMEGA Table
445.3
(17.53)
All dimensions are in mm (inches)
Flat washer
336
(13.23)