GE Medical Systems
REV 5
Illustration 16 – CABLE CURVATURE
THRU–FLOOR
TABLE SIDE
Note:
In case of thru–floor cabling, if the electrical cables are coming from the head side, they will
need to have a minimum curvature with a minimum radius of 150 mm (6").
In any other cases (i.e. flush floor) no such curvature is allowed.
ON
GRADE
4 water pipes between
and LCA H1C positioner
FLOOR
Note:
In case of on grade cabling, because of the minimum curvature constraint of 150 mm (6"),
the cable will have to come from the side between Innova LC Positioner and patient table.
THRU–FLOOR FITTING
(FLOOR)
Chillers
Innova 2000 Cardiovascular Imaging System
ISOCENTER
Electrical cables coming
from positioner cabinet
LCA Baseplate
Electrical cables
coming from cabinets
48
Pre–Installation Manual
pim 2337741–100
WATER LINES
LCA BASEPLATE
Mini radius = 150 mm (6 inches)
(FLOOR)
Cable conduit
(flush floor
interface fitting)
TABLE
SIDE