Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Magstim 200
2
is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment
should assure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
!EC 61000-4-3
IEC 60601
Test Level
3 Vrms
150 kHz to
80 MHz
3 Vim
80 MHz to
2,5 GHz
Compliance
Level
3 Vrms
150 kHz to
80 MHz
3V/m
80 MHz to
2,5 GHz
NOTE
1
At 80 MHz and 800 MHz, the higher frequency range applies.
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the Magstim 200
2
,
including
cables, than the recommended separation distance calculated
from the equation applicable to the frequency transmitter.
Recommended separation distance
3,5
d = [
-----t"'
p--
V1
3,5
_ _ 80 MHz to 800 MHz
d
= [
-----t'i
p
E1
7
800 MHz to 2,5 GHz
d = [
-----t"'
p--
E1
Where
P
is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field Strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a
should be less than the
compliance level in each frequency range_t
Interference may occur in the vicinity of equipment marked with
the following symbol:
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radios. amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured filed strength in the
location in which the Magstim 200
2
is used exceeds the applicable RF compliance level above, the Magstim 200
2
should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
the Magstim 200
2
.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] Vim.
© The Magstim Company Limited
22
MOP01-EN-01