MAGSTIM 2002 Operating Manual page 9

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

0
t
,< . . . ;'.: >,..
.. · .
! ,..
\
~
· · · · ·
..
:,
.J
/,/
3.1.11
Output Control (12)
Magstim 200
2
Operating Manual
The output control allows the power level displayed on the power output display to be adjusted.
Rotating the control clockwise increases the power level. Rotating anti-clockwise decreases the
power level. Each complete 360° rotation changes the output level by approximately 20%.
3.1.12
TriggerButton(13)
The Magstim 200
2
can be triggered to deliver the stimulus by pressing the yellow trigger button along
with one of the trigger buttons located on the coil. The instrument will continuously charge and
discharge if the trigger buttons are pressed and held.
3.1.13
System Stop Button (14)
The Magstim 200
2
can be disarmed by pushing the red stop button. In this mode the disarmed
indicator will illuminate and the instrument will discharge internally.
Note: The Magstim 200
2
will disarm if the coil is disconnected, becomes faulty, the surface
temperature of the connected coil has risen above its set limit, the system stop button is activated, the
instrument detects an internal fault, or if the ready symbol has been illuminated for over 1 minute
without the instrument having been triggered.
3.1.14
Coil Output Socket (15)
This socket is used to connect the stimulating coil to the Magstim 200
2
Note: It is normal for the pin indicated on the label to protrude from the coil socket. This is an
intentional design feature.
Figure 3.2: Coil output socket front view
Figure 3.3: Coil output socket angled view
© The Magstim Company Limited
6
MOP01-EN-01