5-1
Type
SET
LASER
Ramco Innovations (800) 280-6933 www.panasonicsensors.com nsales@ramcoi.com
Corrective Actions
If an error or trouble occurs during use of the system or a system failure is
suspected, read this chapter carefully first and carry out the corresponding
corrective actions.
Trouble type
SET
Trouble in sensor head settings.
Trouble in communication control.
COM
Trouble in the measurement method or display of measurement values.
MEAS
Trouble in alarm or error LED indications.
LED
LASER
Trouble in laser emission.
For the details of conditions where measurement value is unfixed,
Conditions When Output Data Become Unfixed".
For problems related to memory saving and terminal settings,
Memory Change".
Problem
The connecting cable is
not connected properly.
The connecting cable is
disconnected.
The sensor head
indicator does not
Power is not supplied to
light up.
the controller.
The sensor head
does not operate.
The operation of the
sensor head is stopped.
The eco mode is set to
"Eco-FULL".
The laser control setting
The laser beam is
is turned OFF and
not emitted.
saved.
Cause
Corrective action
Check the connection between
the sensor head and connection
cable.
Check the wiring between the
connection cable and connector.
Check the connection between
the 24-VDC external power
and sensor head.
Turn ON the sensor head again.
The LED will be lit by operating
any switch. Make eco-mode
settings if necessary.
Save the laser control setting
turned ON, otherwise the system
will start with the laser beam
turned OFF.
refer to "2-5
refer to "3-3-2
Ref.
page
1-2
2-2
2-2
3-6
3-40
3-39
3-43

Advertisement