5 Liter Oxygen Concentrator Instruction Guide
®
WARNING– Read instRuction guide befoRe opeRating this equipment.
CAUTION– Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
made in the usa of u.s. and imported parts
dangeR–no smoKing
Guía de instrucciones del concentrador de oxígeno de 5 litros de DeVilbiss®
ES
adveRtencia– Lea La guía de instRucciones antes de poneR a funcionaR este equipo.
PRECAUCION– La ley federal (EE.UU.) establece que este aparato sólo lo puede vender un médico o por prescripción del mismo.
fabRicado en ee. uu. de partes nacionales e importadas
peLigRo–no fumaR
Guide d'instructions du concentrateur d'oxygène 5 litres DeVilbiss®
FR
aveRtissement– LiRe Le mode d'empLoi avant d'utiLiseR ce dispositif.
ATTENTION– En vertu de la Loi fédérale américaine, la vente de cet appareil n'est autorisée que par un médecin ou sur ordonnance de ce dernier.
fabRiquÉ auX États-unis avec des pièces des États-unis et des pièces importées
dangeR–ne pas fumeR
525 Series
Serie de 525
Série 525

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 525DS and is the answer not in the manual?

Questions and answers

Mary Jo Abdal-Aziz
May 4, 2025

how to clean filter

1 comments:
Mr. Anderson
May 17, 2025

To clean the filter for the DeVilbiss Healthcare 525DS:

1. Remove the air filter from the back of the unit.
2. If used, remove the oxygen outlet connector.
3. Wash the filter in a solution of warm water and dishwashing detergent.
4. Rinse thoroughly with warm tap water.
5. Dry the filter completely with a towel before reinstalling.

Do not operate the unit without the air filter or if the filter is still damp.

This answer is automatically generated

Summary of Contents for DeVilbiss Healthcare 525DS

  • Page 1 525 Series Serie de 525 Série 525 DeVilbiss 5 Liter Oxygen Concentrator Instruction Guide ® WARNING– Read instRuction guide befoRe opeRating this equipment. CAUTION– Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. made in the usa of u.s.
  • Page 2 EN ENGLISH................................EN-2 TABLE OF CONTENTS IEC Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN - 3 Important Safeguards .
  • Page 3 IEC SymBOLS It is mandatory to read and understand the Double Insulated Danger - No smoking near patient or device . operating instructions prior to use . Attention Consult Instruction Guide Catalog Number Use no Oil, Grease or Lubricants Electric Shock Hazard . Cabinet to be Serial Number Do not use near heat or open flames removed by authorized personnel only .
  • Page 4 ImPORTANT PARTS OF yOuR CONCENTRATOR Please take time to familiarize yourself with your DeVilbiss oxygen concentrator before operating . Front view (Figure A) Operating instructions Green Power light – illuminates when your concentrator is operating . Power Switch | = ON O = OFF Flow meter knob Flow meter...
  • Page 5 OPERATING yOuR dEvILBISS OxyGEN CONCENTRATOR dANGER Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing oxygen therapy. Keep the oxygen concentrator at least 5 feet (1.6 m) from hot, sparking objects or naked sources of flame.
  • Page 6 DeVilbiss Healthcare recommends that at least the following procedures be carried out by the manufacturer or a qualified third party between uses by different patients. NOTE–If preventive maintenance is due at this time, these procedures should be carried out in addition to the servicing procedures.
  • Page 7 Approval Body and Safety Standard CAN/CSA-C22 .2 No . 601 .1-M90 EmC Compliance To EN60601-1-2 *NOTE: The OSD performance at 5°C to 40°C, 95% R.H. through voltage range on the 525DS verified at 670m. Specifications subject to change without notice. A-525D EN - 7...
  • Page 8 dEvILBISS GuIdANCE ANd mANuFACTuRER’S dECLARATION wARNING medical Electrical Equipment needs special precautions regarding EmC and needs to be installed and put into service according to the Electromagnetic Compatibility [EmC] information provided in the accompanying documents. Portable and mobile RF Communications Equipment can affect medical Electrical Equipment. The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the configuration in which it will be used.

This manual is also suitable for:

525 series

Table of Contents