Sunrise Medical TRU-TRAC TT-92B Series Operating Instructions Manual page 15

Portable intermittent traction machine

Advertisement

27.
The Release
Phase Display @)
will light while the
OPERATIONAL PHASE
machine is decreasing the tension applied in
intermittent traction mode.
TN RELEASE
28.
The Rest Phase
Display
will light when
the
machine has released to and is resting at the minimum
OPERATIONAL
PHASE
C Ř
:
E
HA:
force indicated by the Min Reset
Knob (6).
It remains
lit for the number of seconds
indicated by the Rest
Timer Knob
in intermittent traction mode.
RECOMMENDATIONS
A. As with the safe use of any motorized equipment, your TRU-TRAC™
should only be used with a
responsible adult in attendance. For that reason, do not attempt to put yourself in traction with this
machine.
B.
In the event of a loss of power to the machine, traction force should only be released by gradually
lengthening the patient harness adjustment
straps. To prevent damage,
NEVER
TURN
MAX
TRACTION
KNOB
©
TO CORD
RELEASE POSITION
©
WHILE TRACTION
FORCE
IS APPLIED TO
BRAIDED TRACTION CORD (2).
C.
If you are having difficulty operating your TRU-TRAC™
after carefully reviewing
this guide, contact
your supplier for assistance.
If they are unable to resolve the problem, consult a Customer Service
Representative at the factory.
D.
DO NOT USE any accessory or other device with your
TRU-TRAC™ which requires the removal, modification,
or replacement of the acorn nuts shown
in the
illustration to the left. Either overtightening of the
original acorn nuts, or overtightening of any
replacement/adaptor nuts can deform the housing
and
interfere with the safe and proper operation of the
moving parts inside the housing of the TRU-TRAC™.
Read
and follow warning on page 1.
NOTE: The TRU-TRAC
Brand "TD-9000" 3-Dimensional
Cervical System
may be an exception.
If so, follow
accompanying
TD-9000 installation instructions
carefully.
Page 14
No. 990002C (10/85)