CAUTION CONDITIONS
23.
READ AND
FOLLOW
OPERATING
INSTRUCTIONS
PATIENT
INTERRUPT
OPERATIONAL
PHASE
HOLD /
RELEASE
PuuL_7
V REST
OPERATIONAL PHASE
26.
OPERATIONAL PHASE
HOLD /
No. 990002C (10/85)
24.
25.
The yellow colored
Patient Interrupt Display (23)
flashes whenever the Patient Interrupt Switch
9
is
actuated. The light continues to flash after the
Treatment Timer Knob reaches zero to advise the
operator that the patient may
have interrupted the
treatment
prior to completion. A patient interrupted
treatment may be resumed by turning the Power On/Off
Switch (7) "OFF" and "ON" again.
The Operational Phase Section
of the instrument
panel features four orange colored displays that
sequentially light during a treatment to advise the
operator of the current operational
phase status.
The Pull Phase Display (25) will light while the
machine
is increasing the tension applied either in
static or intermittent traction modes.
The Hold Phase Display
will light when the
machine has pulled to and is holding at the maximum
force indicated by the Max Traction Knob ©.
It either
remains
lit for the number of seconds
indicated by the
Hold Timer Knob
in intermittent traction mode, or
for the length of time indicated by the Treatment Timer
Knob
(17) in static traction mode.