You can use the optional auxiliary power lines (Aux1 and Aux2) to draw power for
non-FieldDAQ devices in your network. The FieldDAQ device does not make use of
the auxiliary power itself. If you are connecting power through the auxiliary power
lines, ensure that the total combined current between V Input and Aux power lines
does not exceed 10 A.
Refer to the FD-11601 Specifications for information about the power connectors,
power requirements, and current limits of your FieldDAQ device.
Cap the power connectors when not in use.
Power Connectors
The following figure shows the pinout of the Power IN connector.
Figure 7. Power Connector Pinout
Pin Number
Wire Color*
1
Brown
2
White
3
Blue
4
Black
5
Gray
* Wire color pertains to M125F power cables sold through NI. Other manufacturers' cable wire
colors may vary.
Table 3. Signal Descriptions
V
Chassis Ground ( )
1
5
2
4
Aux2
3
Aux1
C
Signal
Description
V
Positive voltage line
Aux2
Optional line for powering non-FieldDAQ devices
C
Common. Negative voltage line
Aux1
Optional line for powering non-FieldDAQ devices
Chassis Ground. This terminal is internally connected to the
C terminal.