CALIBRATION PROCEDURE
FD-11601
8-Channel, ±10 V Powered-Sensor Voltage Input Device for FieldDAQ
This document contains information for calibrating the FD-11601. For more information about
calibration, visit ni.com/calibration.
Contents
Software Requirements............................................................................................................. 1
Documentation.......................................................................................................................... 2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................3
Calibration Procedure............................................................................................................... 3
Initial Setup....................................................................................................................... 4
Verification........................................................................................................................4
Adjustment........................................................................................................................ 7
EEPROM Update.............................................................................................................. 8
Reverification....................................................................................................................8
Signal Connector Pinout........................................................................................................... 9
Revision History..................................................................................................................... 10
Software Requirements
Calibrating the FD-11601 requires the installation of NI-DAQmx on the calibration system. NI
recommends using the latest NI-DAQmx driver. Earliest driver support version for calibrating
the FD-11601 is listed in the following table.
Driver
NI-DAQmx
You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports many
programming languages, including LabVIEW, LabWindows
Basic .NET. When you install NI-DAQmx, you only need to install support for the application
software that you intend to use.
Table 1. FD-11601 Driver Support
Earliest Version Support for Device Calibration
19.0
™
™
/CVI
, C/C++, C#, and Visual