National Instruments FD-11601 Calibration Procedure

National Instruments FD-11601 Calibration Procedure

8-channel, ±10 v powered-sensor voltage input device for fielddaq
Hide thumbs Also See for FD-11601:

Advertisement

Quick Links

CALIBRATION PROCEDURE
FD-11601
8-Channel, ±10 V Powered-Sensor Voltage Input Device for FieldDAQ
This document contains information for calibrating the FD-11601. For more information about
calibration, visit ni.com/calibration.
Contents
Software Requirements............................................................................................................. 1
Documentation.......................................................................................................................... 2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................3
Calibration Procedure............................................................................................................... 3
Initial Setup....................................................................................................................... 4
Verification........................................................................................................................4
Adjustment........................................................................................................................ 7
EEPROM Update.............................................................................................................. 8
Reverification....................................................................................................................8
Accuracy Under Calibration Conditions...................................................................................8
Signal Connector Pinout........................................................................................................... 9
Revision History..................................................................................................................... 10
Worldwide Support and Services............................................................................................ 10

Software Requirements

Calibrating the FD-11601 requires the installation of NI-DAQmx on the calibration system. NI
recommends using the latest NI-DAQmx driver. Earliest driver support version for calibrating
the FD-11601 is listed in the following table.
Driver
NI-DAQmx
You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports many
programming languages, including LabVIEW, LabWindows
Basic .NET. When you install NI-DAQmx, you only need to install support for the application
software that you intend to use.
Table 1. FD-11601 Driver Support
Earliest Version Support for Device Calibration
19.0
/CVI
, C/C++, C#, and Visual

Advertisement

Table of Contents
loading

Summary of Contents for National Instruments FD-11601

  • Page 1: Table Of Contents

    Worldwide Support and Services.................... 10 Software Requirements Calibrating the FD-11601 requires the installation of NI-DAQmx on the calibration system. NI recommends using the latest NI-DAQmx driver. Earliest driver support version for calibrating the FD-11601 is listed in the following table.
  • Page 2: Documentation

    NI-DAQmx C Reference Help—Provides reference information for NI-DAQmx C functions and NI-DAQmx C properties. Test Equipment The following table lists the equipment required for calibrating your FD-11601. If you do not have the recommended instruments, use the minimum requirements to select substitute equipment.
  • Page 3: Test Conditions

    22 AWG signal wires Test Conditions The following setup and environmental conditions are required to ensure the FD-11601 meets published specifications: • Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
  • Page 4: Initial Setup

    Initial Setup Complete the following steps to set up the FieldDAQ device. Install the software and NI-DAQmx driver as outlined in the FD-11601 Quick Start. Note You must install NI-DAQmx 19.0 or later for device calibration support. Set up the FieldDAQ device as outlined in the FD-11601 Quick Start.
  • Page 5 0.00105 Positive Full Scale 10.0 9.99525 10.00475 The test limits in this table are derived using the values listed in Accuracy Under Calibration Conditions. Set the calibrator to Operate mode (OPR). FD-11601 Calibration Procedure | © National Instruments | 5...
  • Page 6 Repeat steps 3 through 6 for each test point. Set the calibrator to Standby mode (STBY). Disconnect the calibrator from the FieldDAQ device. 10. Connect the calibrator to Bank 2 of the FD-11601, as shown in the following figure. Refer to Signal Connector Pinout for signal connection information.
  • Page 7: Adjustment

    Repeat steps d through e for each calibration voltage in the array. Set the calibrator to Standby mode (STBY). Disconnect the calibrator from the FieldDAQ device. Connect the calibrator to Bank 2 of the FD-11601, as shown in the following figure. Refer to Signal Connector Pinout for signal connection information.
  • Page 8: Eeprom Update

    Worldwide Support and Services for assistance in returning the device to NI. Accuracy Under Calibration Conditions The values in the following table are based on calibrated scaling coefficients, which are stored in the onboard EEPROM. 8 | ni.com | FD-11601 Calibration Procedure...
  • Page 9: Signal Connector Pinout

    FD-11601. For operational specifications, refer to the most recent FD-11601 Specifications at ni.com/manuals. Signal Connector Pinout Figure 6.
  • Page 10: Revision History

    NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the file on your media, or the National Instruments Patent Notice at . You can find patents.txt ni.com/patents...

Table of Contents