Lift covers up slowly to its end position.
!
CAUTION!
The unpacking position shall be shall be clean, level and commodious.
The unpacking personnel shall be well trained, have read and understand fully
relevant instructions in advance.
Report to responsible person immediately if any damages or missing parts.

Mounting the microscope

6.2.1

Recommended table specification

First of all, the recommended specification of the table for mounting the microscope shall meet the
following value requirements in the table:
Designation
Length
Width
Height
Thickness
Weight
Bearing capacity
Distance between feet center and the
surface edge in the projection view
In addition, the following dimension requirements of the table clamp shall be also considered so as to
choose a proper table to mount the microscope.
000000-2466-133 -1.1-2022-06-22 for USA
Risk of damaging instruments in event of knocking down the
cover accidentally!
Value
≥1500mm
≥750mm
730~770mm
20mm-40mm (for stationary version)
20mm-70mm (for mobile version)
≥60Kg
The table shall withstand 4 times the weight of the
device configuration.
<50mm
Inspect all the parts for any transport damage.
Check for any missing parts against delivery
list.
Preparing system for use