MODEL 9100
SOURCE ASSEMBLY REPLACEMENT GUIDE
PUBLICATION N
812106
O
LANGUAGE: ENGLISH
Model 9100 Source Assembly
i
Q U A L I T Y C U S T O M E R S O L U T I O N S

Advertisement

loading

Summary of Contents for Ametek LAND 9100

  • Page 1 MODEL 9100 SOURCE ASSEMBLY REPLACEMENT GUIDE PUBLICATION N 812106 LANGUAGE: ENGLISH Model 9100 Source Assembly Q U A L I T Y C U S T O M E R S O L U T I O N S...
  • Page 2: Health And Safety Information

    IMPORTANT INFORMATION - PLEASE READ Health and Safety Information Read all of the instructions in this booklet - including all the WARNINGS and CAUTIONS - before using this product. If there is any instruction which you do not understand, DO NOT USE THE PRODUCT.
  • Page 3 Web: www.ametek-land.com For further details on all AMETEK Land offices, distributors and representatives, please visit our website. Return of Damaged Goods IMPORTANT If any item has been damaged in transit, this should be reported to the carrier and to the supplier immediately.
  • Page 4 Source Assembly Replacement Model 9100 Changing the Model 9100 Source Assembly Always use a grounded workstation when working on exposed electronic components and take appropriate precautions to avoid damage from electrostatic discharge. Switch off power to the receiver. Disconnect all cables. Refer to Fig. 1. Use a 12 mm wrench to undo the four nuts (A) and remove the transmitter from the purge.
  • Page 5 Source Assembly Replacement Model 9100 Refer to Fig. 2. Remove the power and thermocouple connectors (C). Undo the four slotted screws and two hex nuts (D) then gently remove the old source. Fit the new source. Tighten the four slotted screws and two hex nuts (D).
  • Page 6 Source Assembly Replacement Model 9100 The following step can be omitted if the instrument serial number is 195391 or higher. Refer to Fig. 3. Identify the transceiver processor chip (E). Remove the old processor chip and replace it with the new chip (supplied). Use a specialist chip remover if available;...
  • Page 7 Gas Monitoring Products Standard Terms of Warranty Land Instruments International warrant all products of our own manufacture to be within specified limits of calibration, if any, when despatched from our works. Also that they are free from defects in material and workmanship for normal use and service when used for their intended purpose within the limits of our specification.
  • Page 8 MODEL 9100 SOURCE ASSEMBLY REPLACEMENT AMETEK Land China Service AMETEK Land India Service Land Instruments International AMETEK Land - Americas Dronfield, S18 1DJ, 150 Freeport Road, Pittsburgh, Part A, 1st Floor No. 460 Divyasree N R Enclave,Block A, North Fute Road,...