North American Dräger Narkomed 4 Anesthesia System
North American Dräger Narkomed 4 Anesthesia System
Note
112
Device Driver Name:
Device Driver P/N:
Supported Devices:
Connection:
Connect VueLink to Port A only on Narkomed 4. Otherwise, incorrect data may
be displayed.
If the patient monitor RS-232 MECIF interface is being used, then configure
MECIF only for the highest (in letter, e.g., C is higher than B) used port on the
Narkomed 4. Otherwise, incorrect data may be displayed.
If all of the following conditions are true:
The patient monitor RS-232 MECIF interface is connected to the Narkomed 4.
Another Type B VueLink module (Ventilator, Gas Analyzer, Anesthesia
Machine or Auxiliary Plus) is being used.
The above VueLink module is measuring one or more numerics which are also
output by the Narkomed VueLink interface. (For example, the Philips M1025A/
B Anesthetic Gas Analyzer numerics include !imCO2, !etCO2, !inN2O, etN2O,
!FIO2, !inAGT, !etAGT, !PULSE, !SpO2 and !AWRR. These correspond to the
Narkomed VueLink interface numerics: IMCO2, ETCO2, inN2O, etN2O,
FIO2, inAGT, etAGT, PULSE, SpO2 and AWRR.)
Then the following must be performed to avoid the display of incorrect data:
Each duplicate numeric (from the third condition) should NOT be selected on
NAD NARKOMED 4
M1032-B38rl
where:
r = revision
l = language
Narkomed 4 Anesthesia System Software
Revision 1.30
RS-232 Digital
Protocol:
Vitalink
Word Length:
8 bits
Stop Bits:
1
Parity:
None
Baud Rate:
19200