North American Dräger Narkomed 4 Anesthesia System
North American Dräger Narkomed 4 Anesthesia System
Note
112
Device Driver Name:
Device Driver P/N:
Supported Devices:
Connection:
•
Connect VueLink to Port A only on Narkomed 4. Otherwise, incorrect data may
be displayed.
•
If the patient monitor RS-232 MECIF interface is being used, then configure
MECIF only for the highest (in letter, e.g., C is higher than B) used port on the
Narkomed 4. Otherwise, incorrect data may be displayed.
•
If all of the following conditions are true:
•
The patient monitor RS-232 MECIF interface is connected to the Narkomed 4.
•
Another Type B VueLink module (Ventilator, Gas Analyzer, Anesthesia
Machine or Auxiliary Plus) is being used.
•
The above VueLink module is measuring one or more numerics which are also
output by the Narkomed VueLink interface. (For example, the Philips M1025A/
B Anesthetic Gas Analyzer numerics include !imCO2, !etCO2, !inN2O, etN2O,
!FIO2, !inAGT, !etAGT, !PULSE, !SpO2 and !AWRR. These correspond to the
Narkomed VueLink interface numerics: IMCO2, ETCO2, inN2O, etN2O,
FIO2, inAGT, etAGT, PULSE, SpO2 and AWRR.)
•
Then the following must be performed to avoid the display of incorrect data:
Each duplicate numeric (from the third condition) should NOT be selected on
NAD NARKOMED 4
M1032-B38rl
where:
r = revision
l = language
Narkomed 4 Anesthesia System Software
Revision 1.30
RS-232 Digital
Protocol:
Vitalink
Word Length:
8 bits
Stop Bits:
1
Parity:
None
Baud Rate:
19200
Anesthesia Machines