Engström ELSA/EAS Anesthesia System
172
The configurations for each wave and numeric channel are provided in the following
table.
Channel
Label
Wave 1
CO2CC
Wave 2
AWP
Numeric 1
O2CC
Numeric 2
MIVOL
Numeric 3
FRESH
Numeric 4
-
1. You can configure the number of decimal places by moving the decimal point, using the
left/right arrow keys. The decimal representation in the table above refers to the Value
Range. Please ensure that they are configured as shown.
Value Range =
Unit
Voltage Range
mmHg
0 to 76 mmHg = 0 to 1 V
cmH
O
0 to 100 cmH
2
%
0 to 100% = 0 to 1 V
l/min
0 to 50 1/min = 0 to 1 V
l/min
0 to 50 1/min = 0 to 1 V
-
-
Decimal
Represent.
x
x
O = 0 to 1 V
2
x
x.x
x.x
-
Free Analog Device Information
1

Advertisement