Philips VueLink M1032A Service Booklet page 175

Table of Contents

Advertisement

Step 1: Ensure that all the necessary hardware is available
Note
Step 2: Prepare the free analog cable
Free Analog Device Information
General information on configuring the free analog devices has not been included
in this technical note. Please refer to Chapter 3 of the VueLink Handbook, if you are
not familiar with this procedure.
(10 Minutes)
You will need:-
An M1032A #A05 VueLink Module
4m Free Analog (open end) cable M1032A #K6A or M1182A #K6A
or
2m Free Analog cable M1032A #K60 or M1182A #K60.
Seven plugs to fit the Nellcor analog output ports:
7 x 3/32 inch subminiature phone plugs (single tip-mono)
Seven cables to connect the plugs to the open end of the Free Analog cable:
7 x shielded cable (length: 10 - 15cm) (diameter: 2.5mm, RG145 or similar)
These cables and plugs are provided locally by the hospital or by the manufacturer;
they will not be set up at SMO/SLI.
The connector hood, the divider (5 holes) and five sleeves to connect the
seven cables to the Free Analog cable are included with the free analog cable.
(75 Minutes)
1. Connect each of the seven short cables (cable 1 to 7) to a phone plug.
2. Put all of the included sleeves through the holes of the divider (A
Nellcor N1000 and N2500 Multi-Function Monitor
to A
).
1
5
175