Step 4: Configure each wave and numeric channels
164
(30 Minutes)
The configurations for each wave and numeric channel are provided in the following
table.
Channel
Label
Wave 1
AWP
Wave 2
!expFl
Numeric 1
!MIVOL
Numeric 2
!AWRR
Numeric 3
!TV
Numeric 4
!FIO
!COMP
3
Numeric 5
!Rstat
1. You can configure the number of decimal places by moving the decimal point on the input
value range, using the left/right arrow keys. The decimal representation in the table above
refers to the Value Range. Please ensure that they are configured as shown.
2. Evita only provides positive voltages for expFl (Expiratory Flow).
3. Only 1 available at a time, depending on configuration.
Unit
mbar
-100 to 0 mbar-10 to 0 V
0 to 100 mbar0 to 10 V
l/min
2
0 to 150 l/min0 to 10 V
l/min
0 to 35 l/min0 to 10 V
1/min
0 to 60 1/min0 to 10 V
l
0 to 2 l0 to 10 V
none
0.00 to 1.000 to 10 V
2
ml/mbar
0 to 255 ml/mbar0 to 10V
mbar/L/s
0 to 100 mbar/L/s0 to 10V
Value Range =
Voltage Range
Free Analog Device Information
1
Decimal
Represent.
x
x
x.x
x
x.xx
x.xx
x
x