Braun Fm Device Information - Philips VueLink M1032A Service Booklet

Table of Contents

Advertisement

Braun FM Device Information

Braun FM Device Information
Caution
134
Device Driver Name:
Device Driver P/N:
Supported Devices:
Connection:
Please upgrade the IntelliVue Software to Rev. C.00.90 (patch) or higher or
IntelliVue Rev. D.0 to avoid any loss of VueLink data from the resting
display.
Please refer to the respective IntelliVue FCO and Service Bulletin.
Explanation: If you change the VueLink labels on the main screen, labels
and values get mixed up. The values appear with the wrong label.
The values in the VueLink task window are correct.
Braun fm-system
M1032-B33rl
where:
r = revision
l = language
Braun fm system with fm computer
Software Revision 3.10
RS-232
Baud Rate:
9600
Word Length:
8 bits
Stop Bits:
1
Parity:
None