116
Device Driver Name:
Device Driver P/N:
Supported Devices:
Connection:
The word length, start and stop bits, parity and baud rate must be set and verified
according to the NAD Narkomed 6000 Technical Service Manual
1. Connect VueLink to the one of the RS-232 ports COM5 to COM8 on
Narkomed 6000.
NAD NARKOMED 6000
M1032-B49rl
where:
r = revision
l = language
North American Dräger Narkomed 6000
Anesthesia System
Software Revision 1.06
RS-232 Digital (use one of the ports: com5 -
com8)
Protocol:
Vitalink (fixed)
Word Length:
8 bits
Start Bits:
1
Stop Bits:
1
Parity:
None
Baud Rate:
19200
Anesthesia Machines