Radiometer Transcutaneous pO2/pCO2 Monitor
Step 2: Select two Numerics
Step 3: Prepare the free analog cable
184
(10 Minutes)
The following parameters are provided at the Analog Output port of TCM3:
Alarm (TCM3 pin 5) *Transducer Temperature (TCM3 pin 8)
Transcutaneous partial oxygen pressure (TCM3 pin 9)
Transcutaneous partial CO
In the following, the configuration is described using a typical selection of the two
numerics and the alarm:
Transcutaneous partial oxygen pressure
Transcutaneous partial CO
(45 Minutes)
1. Connect the black and white wires that are twisted together and the thick
black wire to the TCM3 Analog ground (pin 15). Do NOT separate these
wires. Wire pin 13 and pin 15 together.
2. Connect the colored wires to the analog output connector on the external
device. The respective channels for each wire color are given in the following
table.
VueLink Free Analog Cable
Channel
---
Black and White (twisted)
---
Black (thick)
Bridge
pressure (TCM3 pin 10)
2
pressure
2
Wire Colours
Designation
TCM3 PIN
Analog
15
ground
Shield
153
13 to 15
Free Analog Device Information