4. Assembly
I
· • 6. Eyepiece tube assembly
1 Eyepiece Tube
Completely loosen the eyepiece tube clamp screw on the imermediate tube with the
hexagonal screwdriveL Fit the circular dovetail of the eyepiece tube into the circular
dovetail groove of the intermediate tube by tilting it at an angle. When filling. insert the
positioning pin on the eyepiece tube in the receiving groove on the intem1ediate tube. Lock
it in position by tightening the clamp screw.
Remove any looseness between the positioning pin and groove by pushing in the eyepiece
tube while rotating in the clockwise direction.
2 Eyepieces
Attach eyepieces of the same magnification. Attach the eyepiece containing crosshairs on
the right side so as to be viewed with the right eye. The sleeve has a positioning pins.
Insert the eyepieces by aligning the grooves of the eyepiece with the positioning pins of the
eyepiece tube sleeve. Insert the rubber eyegard (sold separately) so that they fit into the
grooves around the outside of the eyepieces.
7. Power cord
·;,
.
. ' Ii\
WARNING
Use only the following power supply cord set.
•For I00--120V area
UL Listed. detachable cord set. 3 conductor grounding type SVT. No. 18
AWG rated at 125V. 7A minimum.
In case of using the extension cord. use only the power supply cord with the
PE (protective earth) wire.
• For 220--240V area
3 pole power supply cord set. which must be approved according to EU/EN -
standards.
Class I equipment should be connected
to
PE (protective earth) terminal.
In case of using the extension cord. use only the power supply cord with the
PE (protective earth) wire.
Make sure to turn off the microscope power switch (flip it to the
"0''
side) before connecting the
power cord.
Plug the cord into the socket of the AC input connector on the back of the microscope. Securely
plug the other end of the cord into an AC outlet.
Covering the tapped holes with protective stickers
The four tapped holes on the upper surface of the microscope arm are for mounting accessories.
such as Epi-fluon:scence attachment. etc. When not used, cover the tapped holes with the provided
protective stickers.