Nikon ECLIPSE E600 POL Instructions Manual page 34

Polarizing microscope
Hide thumbs Also See for ECLIPSE E600 POL:
Table of Contents

Advertisement

34
1. Confirming the input voltage
i::J.
-
· -
ltllton
MOOEL ECLIPSE E600 POL
100-120V-
24A
SOISOHZ
430001
Make sure that the input voltage indicated on the back panel of the microscope
is the same as the voltage provided in your area. If the indicated voltage is
different. do not use the microscope and notify your nearest Nikon
representative immediately. If the microscope is used with the wrong input
voltage. a short circuit. electrical shock or fire may result. causing damage to
the microscope.
2. Lamp and lamphouse (lamp replacement)
CAUTION
• To prevent electrical shock and damage to the microscope. always turn off the
...._.._,.
power switch (flip it to the
"0"
side) and unplug the power cord when
connecting or disconnecting the lamphouse.
• To prevent burns. allow the lamp and lamphouse to cool before replacing it
for at least 30 minutes after using.
• Use the Nikon C-LP HALOGEN 12V-IOOW model for the lamphouse.
•Use a 12 V-100 W LONGLIFE halogen lamp (OSRAM HLX 64623 or
PHILIPS 7724 ).
• Do not touch the glass surface of the lamp with your bare hands. Doing so
will cause fingerprints, grease. etc .. to burn onto the lamp surface, reducing.
the illumination provided by the lamp. If you do get any fingerprints or dirt
on the lamp. wipe it clean.
• After replacing the lamp. make sure that the lamphouse cover is attached
securely. Never use the lamphouse while its cover is off.