6.5 Calibration
Note: Calibration must be done in the following three occasions:
1. Before using new sensors or new meter.
2. When the sensor been used for a long time and cause a wide measuring result.
3. When reference temperature or temperature factor need to be changed.
Calibration step
dal"
ru
DD
LI, Zt LI
1. In measurement mode, press
to enter calibration mode.
2. Dip the electrode into standard solution.
a
|
3. After you get a stable current value or the
is snowed, press 9
to adjust the value
figures to match with the standard solution. Adjustable range is +30% of the stable current value.
Press
to finish the calibration process.
4. The value of cell constant K indicates the status of the electrode.