Immunity to proximity fields from RF wireless communication equipment

7.2
The portable or mobile RF communication devices that may be present in the home, such as wireless
telephones, mobile phones, devices for wireless connection to the Internet and similar, must be kept
at a distance from BIOSTIM SPT, to avoid the risk of interference. The recommended minimum
separation distance depends on the output power of the RF device and the transmission frequency.
The user can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications systems and BIOSTIM SPT, using reference the table
below.
Test
Band
Frequency
(MHz)
(MHz)
385
380 –390
450
430 – 470
710
745
704 – 787
780
810
870
800 – 960
930
1 720
1 845
1700 –1990
1 970
2 450
2400 –2570
5 240
5 500
5100 –5800
5 785
The immunity levels shown in the table are respected as long as the device is maintained at a
distance of at least 30 cm from any possible source of RF disturbance.
BIOSTIM SPT, complies with all test levels with a distance >= 30 cm.
Note: BIOSTIM SPT, at the frequency of 450 Mhz is immune to the disturbance up to a level of 20V /
m - To obtain the required immunity level (28 V / m) the distance from the source of disturbance must
be increased to not less than 42cm.
8.

SYMBOLS

Class II appliance: Appliance in which protection against electric shock does not
rely on basic insulation only, but includes additional safety precautions such as
double insulation.
Appliance with BF type applied part: appliance with a specific degree of protection
against electrical hazards, specifically regards admissible leakage current, and with
an F type (floating) applied part isolated from the rest of the appliance (coil).
Continuous current: on the generator data plate indicates the required power
supply, on the power supply data plate indicates the provided power supply.
Alternate Current (AC): on the power supply data plate indicates the required power
supply.
24 24
BIOSTIM SPT – User Manual
Type of service
Modulation
Pulse modulation
TETRA 400
GMRS 460, FRS
± 5 kHz deviation
460
1 kHz sine
LTE Band 13,
Pulse modulation
17
217 Hz
GSM 800/900,
TETRA 800, iDEN
Pulse modulation
820, CDMA 850,
LTE Band 5
GSM 1800; CDMA
1900; GSM 1900;
Pulse modulation
DECT;
217 Hz
LTE Band 1, 3,
4, 25; UMTS
Bluetooth, W LAN,
802.11 b/g/n, RFID
Pulse modulation
2450,
217 Hz
LTE Band 7
Pulse modulation
W LAN 802.11 a/n
217 Hz
BIOSTIM SPT User Manual-EN
Maximum
power
(W )
1,8
18 Hz
FM
2
0,2
2
18 Hz
2
2
0,2
Distance
Immunity Test Level
(m)
(V/m)
0,3
27
0,3
28
0,3
9
0,3
28
0,3
28
0,3
28
0,3
9