Maintenance Trouble Shoot; Technical Specifications - Proactive Protekt Aire 2000 Operation Manual

Table of Contents

Advertisement

MAINTENANCE TROUBLE SHOOT / TECHNICAL SPECIFICATIONS
Trouble Signal
No air output
Incomplete inflation
PACKAGE CONTENT LIST
• Control Unit x 1

TECHNICAL SPECIFICATIONS

INPUT RATING: 120VAC, 60Hz, 1A FUSE RATING: 250V, 1A
Control Unit
Item #: 80021
Power Supply: 120V/60Hz
Air output: 8 liter/min
Pressure range: 30-60 mmHg
Cycle Time: 10 min
Size: 11'' x 5.9'' x 3.5''
Weight: 1.4kg (3lbs)
Power indicator
is ON.
Power indicator
is OFF.
overlay
too soft.
• Overlay Unit x 1
Failure Cause
The inflation tube is not con-
nected to the control unit.
The silicone tubes inside the
control unit are disconnected.
No power.
Small hole at the air cell, or
overlay life comes to its end.
Air knob indicate at "MIN"
location.
• User Manual x 1
Top Cover Material: Nylon/PU
Air Cells Material: Nylon/PVC
Bottom cover material: Nylon/PVC
9
Fault Handling
Connect the air tube.
Contact Service Personnel.
Turn on power switch
Replace the air cell or
the whole overlay.
Turn the air tube at the
right to give max pressure .
Overlay
Item #: 80022
Size: 80''L x 36''W x 5''H
# of aircells: 20
CPR Valve? YES