Maintenance And Inspection - Proactive PROTEKT STS COMPACT 500 Manual

Sit-to-stand patient lif t
Table of Contents

Advertisement

Maintenance & Inspection
• The Protekt
STS Compact 500 requires a minimum of service to keep it in proper
®
working order and to assure safe operation. It is important that certain basic checks be
performed periodically by the operator of the lift or by a qualified bio-med or
maintenance staff.
• The manufacturer recommends that the following components and operating points be
scheduled for periodic inspection at intervals not greater than 6 months (≤ 6). The
manufacturer also recommends visual inspection prior to use for the additional points
Maintenance & Inspection
below (V). The periodic inspection interval and the visual inspection points are listed
below.
The Bestcare Lift requires a minimum of service to keep it in proper working order and to assure safe operation. It
• Note – Proactive assumes that the assembly instructions were performed properly and
is important that certain basic checks be performed periodically by the operator of the lift or by a qualified bio-med
or maintenance staff.
by a qualified individual. Proactive also assumes that the operator of the lift has been
The manufacturer recommends that the following components and operating points be scheduled for periodic
properly trained, reviewed all the training material and knows how to operate the lift.
inspection at intervals not greater than 6 months (≤ 6). The
• With proper use and care, the expected lifetime of the lift is 10 years or 20,000 cycles.
to use for the additional points below (V). The periodic inspection interval and the visual inspection points are
listed below.
• The expected lifetime of the electrical components are 3 years.
Note – Bestcare assumes that the assembly instructions were performed properly and by a qualified individual.
• Contact Proactive for any recycling infomation.
Bestcare also assumes that the operator of the lift has been properly trained, reviewed all the training material and
knows how to operate the lift.
Boom and Spreader Bar
Check connections between boom and spreader bar for proper connection with pin
& key ring (for Patient Lifts)
Check connection between boom and mast for improper connection, looseness, or wear
Check spreader bar hook bolts for looseness or damage (for Patient Lifts)
Key rings are present on spreader bar hooks (on patient lifts) and on boom hooks
(on stand assist lifts)
Mast
Check mast for bending or deflection
Base and Foot Pedal
Check all nuts and bolts for looseness
Check all casters and axle bolts for tightness
Check all casters for rotation and locking/unlocking (rear only)
Apply multi-purpose grease to bearings if needed
Check all weld joints for rust and cracks
Actuator and Control Box
Confirm control box is securely affixed to the mast, and battery pack is secured in control
box
Confirm actuator is secured between the boom and the mast with the pins and key
rings
Confirm manual emergency lowering mechanism on actuator operates properly
Confirm red button on control box shuts off all power when pressed
Confirm all cables at the bottom of the control box are properly connected
Sling
Inspect the body of the sling and all loops or connectors that could attach to the
spreader bar for signs of wear, damage, or fraying
Lift Cleaning
manufacturer
also recommends visual inspection prior
23
V
≤ 6
upon use
months
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
As
As
needed
needed

Advertisement

Table of Contents
loading

This manual is also suitable for:

Protekt sts50034500-com