CLEANING GUIDELINES
• Use a neutral detergent to clean surface of control unit and overlay; disinfectant products may be
used according to the manufacturer's protocol. Do not use phenolics on this product.
• Do not immerse the control unit in water.
• Do not heat or steam in autoclave.
• Be sure to fully air dry the overlay and cover after cleaning and before use.
• After using for some time (~ 3 months), clean the air filter cotton inside the enclosure base, steps as
below:
1) Take out air filter cover and air filter cotton.
2) Wash air filter cotton with clean water, if the dirt sticks to the filter, soak the air filter cotton in the
water.
3) Dry the air filter cotton then put it back to the air filter cover.

MAINTENANCE

• Make sure the control unit is in good condition by checking if the indicators illuminate when the
power is first turned on.
• Power supply cord can be replaced by SERVICE PERSONNEL.
Power supply Information: Non-detachable cord , 18AWG/2C, Rated 300V, 105℃.
!
WARNING
For safety reasons, only qualified service personnel should open the equipment.
MANUFACTURER will provide circuit diagrams, component part lists, descriptions, calibration
instructions to assist to SERVICE PERSONNEL in parts repair.
LAY OPERATOR or LAY RESPONSIBLE ORGANIZATION should contact the MANUFACTURER or the
MANUFACTURER'S representative for assistance in setting up, using or maintaining the product and to
report unexpected operation or events.