OPERATING INSTRUCTIONS

Step 1
Turn on the control unit power. The indicator of the power switch will come on. The control unit will start
to blow air into the overlay.
Step 2
The Low Pressure indicator (orange) will come on as inflation is underway.
Step 3
Press the Static button for a quicker inflation.
Step 4
When the pressure reaches the preset level, within approximately 30 minutes, the Normal Pressure
indicator will come on, and the Low Pressure indicator will turn off.
Step 5
Patients can directly lie on the overlay or cover with a sheet and tuck loosely to increase the comfort of
the patient.
Step 6
Determine the patient's weight and set the control knob to that weight setting on the control unit.
Step 7
Press the Static button to shift between Alternating mode and Static Mode. When in Static mode, and
the Static indicator will come on. The static mode will be started within approximately 6 minutes. In
Alternating Pressure mode, the air cells will alternate in 10 min cycles.
NOTE!
In static mode, the overlay provides a firm surface that makes it easier for the patient to transfer or
reposition. The static mode will help ensure the patient does not bottom out when in a sitting position.
DISCONNECT DEVICE
To fully disengage the power to the unit, disconnect the power cord from the AC inlet.
!
WARNING
• The control unit provided should only be used with the air overlay provided.
• Do not expose the product to lint, dust and sun light to prevent damage of the product.
• Keep the product away from the heat and moisture.
• Keep the product away from pets and children.
ENVIRONMENT REQUIREMENTS
• Operation Temperature: 50℉~95℉ (10℃~35℃)
• Storage and Transport Temperature: 5℉~122℉ (-15℃~50℃)
• Operation Humidity: 20%~80% non-condensing