Keys
• Turns the device on or off
• In treatment mode, press to
stop treatment
• Use this key to exit memory
or setting mode
• In user select mode, press
to select a user (U1 or U2)
• In memory mode, press to
check data or confirm its
deletion
• In treatment mode, use to
pause or resume therapy
• In standby, press to select a
therapy mode
• In standby, press P to select
a therapy program
• Press and hold P to enter
setting mode
• Press to confirm user
selection
• In standby, press to select
a treatment site or press
and hold for 3 seconds to
enter memory mode
• In memory mode, press
and hold for 3 seconds to
clear memory data
• Increases / decreases the
output intensity of channel 1
• In setting mode, press Ch1-
to select the next parameter
• Increases / decreases the
output intensity of channel 2
• In setting mode, press Ch2+
or Ch2- to ajust user
program parameters
8

INSTRUCTIONS FOR USE

1. Prepare the Unit
Before use, charge Thera3+ with the
enclosed USB cable. It can be plugged
into a computer's USB port or into a wall
socket using a adaptor (not included).
When battery is fully charged the LED
indicator will change from yellow to green.
Charging time: 3 hours
2. Connect electrodes to lead wires
Insert the lead wire connector into
the electrodes connector, ensuring
that no bare metal remains exposed.
3. Connect lead wires to the device
This device has 2 output channels,
which can be used simultaneously
(with 2 electrode pairs) or individually
(1 channel with 1 electrode pair). After
making sure that the device is turned
off, insert the wire plugs into the sockets
at the top of the unit.
Caution: Always use the lead wires
supplied with this unit.
4. Place electrodes on skin
Ensure that the skin surface where the electrodes will be placed is
clean and dry. Apply electrodes to the exact site indicated by your
physician. The electrodes should be placed firmly and evenly on the
skin to ensure good contact.
See p.23 for our electrode placement guide, which also includes
recommendations for stimulation positions.