Troubleshooting
EN-
Troubleshooting
US
CAUTION
Any oil leaking from the X-ray tube in the event of a fault is harmful.
Wipe up any oil immediately.
❯
Do not swallow the oil.
❯
Stop using the unit and inform a service technician.
❯
9
Tips for operators and service technicians
Any repairs above and beyond routine maintenance may only be done by suitably qualified per-
sonnel or by one of our service technicians.
Error
Unit does not switch on
Unit not responding
56
Possible cause
EMERGENCY STOP SWITCH
accidentally activated
No mains voltage
On/off switch is defective
The unit has not yet completed
the startup procedure
Cables not correctly connected
Plug-in contacts of the fiber
optic cable are contaminated
Driver for PCI Express frame
grabber card is not installed or
incorrectly installed
COM port incorrectly configured
Remedy
Releasing the EMERGENCY
❯
STOP SWITCH.
Check the mains cable and
❯
electrical connection; replace
if necessary.
Contact technician.
❯
Check the mains fuse in the
❯
building.
Contact technician.
❯
After switching on, wait until
❯
the booting process has fin-
ished.
Check the cable connections.
❯
Clean plug-in contacts and
❯
sockets.
Reinstall the driver or the
❯
complete ProVecta plugin.
Check the COM port in the
❯
service tool.