Electrostatic Discharge (Esd) Precautions - Keysight Technologies 34959A Service Manual

Data acquisition system
Table of Contents

Advertisement

– On the 34945A Microwave Switch/Attenuator Driver, this command performs a
– If you have a 34951A Isolated DAC Module installed, the self-test will
If one or more tests fail, return the instrument to Keysight for service.
Self Test Error Numbers
On the remote interface, a self–test failure will generate SCPI error –330 and a
supplemental message indicating one of the test numbers shown in the table on
page
Calibration Errors
The table on

Electrostatic Discharge (ESD) Precautions

Almost all electrical components can be damaged by electrostatic discharge
(ESD) during handling. Component damage can occur at electrostatic discharge
voltages as low as 50 volts.
The following guidelines will help prevent ESD damage when servicing the
instrument or any electronic device.
– Disassemble instruments only in a static–free work area.
– Use a conductive work area to reduce static charges.
– Use a conductive wrist strap to reduce static charge accumulation.
– Minimize handling.
– Keep replacement parts in original static–free packaging.
– Remove all plastic, foam, vinyl, paper, and other static–generating materials
– Use only anti–static solder suckers.
Keysight 34980A Service Guide
self-test of the 34945A and all connected 34945EXT remote modules.
require an additional 15 seconds to complete per DAC module (a memory test
is performed).
80.
page 84
shows failures that may occur during a calibration.
from the immediate work area.
Obtaining Service
1
33

Advertisement

Table of Contents
loading

This manual is also suitable for:

3262695

Table of Contents