34952A Verification
Analog Output Verification Test
calibration of the analog outputs on the 34952A Multifunction Module.
Verification checks are performed only for those output values with unique
calibration constants.
1 Use the internal DMM to measure the output of each channel. You will need to
2 For each analog output (channel 6 and channel 7), set each output value in the
It is not necessary to test the voltage output at the full rated 10 mA load.
NOTE
Output Voltage
10 V
0 V
-10 V
Keysight 34980A Service Guide
physically move the input connections on the 34952A module to channel 6
and then channel 7.
table on the next page. Compare measurement results to the appropriate test
limits shown in the table.
This procedure is used to check the
Quick Check
Q
Q
Calibration Procedures
Error from Nominal (1 year)
± 45 mV
± 20 mV
± 45 mV
3