SPECIFICATIONS

Model No.: FORA MD Control
Dimension & Weight: 122 (L) x 64 (W) x 24 (H) mm, 130 g
Power Source: One 3.7 V rechargeable Li-ion battery
Display: LCD
Memory: 1000 test results and an additional of 500 sets of control
solution testing data
Auto Shutdown: After idle for 10 minutes
External output: Mini USB cable
Operating Condition: 10°C to 40°C (50°F to 104°F), below 85%
R.H. (non-condensing)
Meter Storage/Transportation Conditions: -20°C to 60°C (-4°F to
140°F), below 95% R.H.
o
o
o
Strip Storage/Transportation Conditions: 2
C to 32
C (35.6
F to
o
89.6
F), below 85% R.H.
Measurement Units: mg/dL or mmol/L
Measurement Range: 20 ~ 600 mg/dL (1.1 ~ 33.3 mmol/L)
Expected Service Life: 5 years
This device has been tested to meet the electrical and safety requirements of:
IEC/EN 61010-1, IEC/EN 61010-2-101, EN 61326-1, IEC/EN 61326-2-6.