4. Press S button to scan the Test Strip ID
barcode or manually enter the Test Strip
ID via the keypad.
5. Press S button to scan Level 1 control
solution lot number or manually enter
the control solution lot number via the
keypad.
6. With the contact bars facing up, insert
the test strip into the test strip slot.
NOTE:
The "Apply Sample" screen will appear.
7. Apply a drop of control solution to the
absorbent hole of the test strip until the
confirmation window is fully filled with
blood.
8. Wait for the meter to finish analyzing
the sample and display the test result.
NOTE:
The meter will produce a beep sound and then begin to count down. The
results will display on the screen and automatically save in the meter.