5. Press S button to scan the Patient ID
barcode or manually enter the Patient
ID via the keypad.
6. Press Up/Down button to select
Capillary/Venous and then press S
button to go to the next step.
NOTE:
Select VENOUS only for venous samples. Select CAPILLARY for all other
samples.
7. Press Up/Down button to select AC/PC,
and then press S button to go to the
next step.
8. With the contact bars facing up, insert
the test strip into the test strip slot.
9. Apply a drop of blood directly from the
patient's finger to the absorbent hole of
the test strip until the confirmation
window is fully filled with blood.