8.5
Function mode switch
8.5.1 Alternate
Alternating mode is the default mode for the system. Within this mode the mattress will
operate in an alternating 1-in-2 cell cycle. The alternating cycle will continue at the
selected cycle time until another mode is selected.
8.5.2 Static
Pressing the Function Mode Selection button until the Static indicator illuminates puts
the system into Static mode.
Within this mode the mattress will maintain a selected constant pressure. After 30
minutes the system will automatically revert back to Alternating mode.
8.5.3 Constant Low Pressure (CLP)
Pressing the Function Mode Selection button until the CLP indicator illuminates puts
the system into Constant Low Pressure (CLP) mode.
Within this mode the internal cell pressures within the mattress will be reduced by 50%
and the mattress will provide a comfortable static surface.
8.6
Auto-Firm
Auto-Firm mode can be selected by pressing the Auto-Firm button. In this mode all cells
inflate to a single pressure setting to provide a firm and stable surface for nursing
procedures or for patient ingress/egress. The system will automatically return to
Alternating mode at the previously selected comfort level after 20 minutes.
8.7
Alternate cycle time selection
The alternating cycle times can be selected to provide an individualised care program for
each patient.
Within Alternating mode the alternating cycle time can be selected by pressing the
Cycle button. Selections can be made from 10-30 minutes at 5 minute intervals. The
cycle time will be displayed in the LCD screen.
8.8
Alarm mute
The Alarm Mute button temporarily resets the audible Low Pressure/Power
Failure/Service alarms. Should the situation not be resolved and the fault condition
continues the alarm will resume notifying carer.
Issue: 5 - 12/12/2018