8.4
Control panel lockout
If the control panel is not used for a period of 30 seconds it will lock out and inhibit the
use of the functions. Additionally, if you wish to lock out the control panel press the
Lock Out button and all functions will be locked. This is to prevent the system being
altered accidently.
In order to unlock the control panel, simply press and hold the Lock Out button for 3
seconds, the control panel will now be active for use.
8.5
Function mode switch
8.5.1 Alternate
Alternating mode is the default mode for the system. Within this mode the mattress
will operate in an alternating 1-in-2 cell cycle. The alternating cycle will continue at the
selected cycle time until another mode is selected.
8.5.2 Static
Pressing the Function Mode Selection button until the Static indicator illuminates puts
the system into Static mode.
Within this mode the cells are not alternating and the internal cell pressure within the
mattress is set at 15mmhg to provide a comfortable static surface.
8.5.3 Constant Low Pressure (CLP)
Pressing the Function Mode Selection button until the CLP indicator illuminates puts
the system into Constant Low Pressure (CLP) mode.
Within this mode the cells are not alternating and the internal cell pressure within the
mattress is set at 15mmhg to provide a comfortable static surface.
8.6
Auto-Firm
Auto-Firm mode can be selected by pressing the Auto-Firm button. In this mode all
cells inflate to a single pressure setting to provide a firm and stable surface for nursing
procedures or for patient ingress/egress. The system will automatically return to
Alternating mode at the previously selected comfort level after 20 minutes.
Issue 9 – 15/09/2022
Page 14 of 28