Operation Below 45 Hz

The following operating characteristics apply for output frequencies between 45Hz and 1Hz. Below 1
Hz, instantaneous values meet the dc specifications. The ac source output is set to: sinewave, dc coupled,
real-time regulation, and is connected to a linear load.
Real Power (watts)
1750
750
375
675
287
142
0
2
4
1
Frequency (linear scale)
RMS current derating (A)
13
6.5
3.75
7
3.5
5 Hz
1.75
0
1
AC coupled frequency response (dB)
0
-10
slope of 1 (20dB/decade)
-20
1
1
.
Table A-3. Operation Below 45 Hz
10 Hz
5 Hz
6
8
45 Hz
10
Frequency
0.1%error @ 40Hz
1%error @ 10Hz
30% error @ 2Hz (-3dB point)
0
1
Frequency
Reactive power (VA)
1750
750
375
0
10
1
2
RMS current programming accuracy (% error)
10
5
2% error at 10 Hz
100
0
10
0
Meets Measurement Accuracy specifications under the following
conditions:
Minimum
measured
Sample rate
frequency
25 s
45 Hz
50 s
22.5 Hz
100 s
11.25 Hz
200 s
5.625 Hz
250 s
4.5 Hz
(Anti aliasing filters have a corner frequency of 12.5 kHz.)
100
Specifications - A
8
4
6
Frequency (linear scale)
20
30
Frequency
Maximum allowable
frequency component
to avoid aliasing
20 kHz
10 kHz
50 kHz
2.5 kHz
2 kHz
10 Hz
10
40
63

Advertisement