Making a Measurement

Making a Measurement with LabVIEW

1.
Launch LabVIEW.
2.
Select Help»Find Example.
3.
Open the example VI that you want to use by selecting Hardware Input and Output»
FlexRIO.
4.
Follow any setup, configuration, and execution instructions in the VI.

Troubleshooting

If an issue persists after you complete a troubleshooting procedure, contact NI technical
support or visit ni.com/support.
What Should I Do if the PXIe-5774 Doesn't Appear in
MAX?
1.
In the MAX configuration tree, expand Devices and Interfaces.
2.
Expand the Chassis tree to see the list of installed hardware, and press <F5> to refresh
the list.
3.
If the module is still not listed, power off the system, ensure that all hardware is correctly
installed, and restart the system.
4.
Navigate to the Device Manager.
Operating System
Windows 10/8.1
Windows 7
5.
Verify the PXIe-5774 appears in the Device Manager.
a)
Under an NI entry, confirm that a PXIe-5774 entry appears.
b)
If error conditions appear, reinstall FlexRIO and the PXIe-5774.

What Should I Do if the PXIe-5774 Fails the Self-Test?

1.
Restart the system.
2.
Launch MAX, and perform the self-test again.
3.
Power off the chassis.
4.
Reinstall the failed module in a different slot.
Description
Right-click the Start button, and select Device Manager.
Select Start»Control Panel»Device Manager.
Note
If you are using a PC with a device for PXI remote control system,
under System Devices, also confirm that no error conditions appear for the
PCI-to-PCI Bridge.
PXIe-5774 Getting Started Guide | © National Instruments | 13