Call With The Following; Parameters: Function = Range = Resolution; Nidmm_Val_Dc_Volts; Call Nidmm_Read . Verify That This Measurement Falls Between The - National Instruments PXI-4070 Manual

Digit flexdmm calibration procedure
Hide thumbs Also See for PXI-4070:
Table of Contents

Advertisement

Table 2. niDMM_ConfigureMeasurement Parameters (Continued)
Calibrator Output
Iteration
Amplitude
5
50 mV
50 mV
6
50 mV
50 mV
7
50 mV
50 mV
© National Instruments Corporation
niDMM_ConfigureMeasurement Parameters
Frequency
50 kHz
NIDMM_VAL_AC_VOLTS
50 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
100 kHz
NIDMM_VAL_AC_VOLTS
100 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
300 kHz
NIDMM_VAL_AC_VOLTS
300 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
14. Output 500 mV at 30 Hz on the calibrator.
15. Call
niDMM_ConfigureMeasurement
parameters:
Function =
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range =
0.5
Resolution = 500
16. Call
niDMM_Read
limits listed in Table 12.
17. Refer to Table 3 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a.
On the calibrator, output the value listed under Calibrator Output
in Table 3 for the current iteration.
b.
Call
niDMM_ConfigureMeasurement
NIDMM_VAL_AC_VOLTS
in Table 3 for the current iteration.
c.
Call
niDMM_Read
limits listed in Table 12.
d.
Call
niDMM_ConfigureMeasurement
NIDMM_VAL_AC_VOLTS_DCCOUPLED
e.
Call
niDMM_Read
limits listed in Table 12.
Function
e–9
. Verify that this measurement falls between the
and the remaining parameters as shown
. Verify that this measurement falls between the
. Verify that this measurement falls between the
21
Range
Resolution
0.05
50e–9
0.05
50e–9
0.05
50e–9
0.05
50e–9
0.05
50e–9
0.05
50e–9
with the following
with Mode set to
again, changing Mode to
.
NI PXI-4070 Calibration Procedure