Gravitational Constant Adjustment Location < Graadj - KERN DS 3K0.01S Operating Instructions Manual

Platform balance
Table of Contents

Advertisement

7.8.3 Gravitational constant adjustment location < graadj >
TDS-A-BA-e-2311
 Press and hold the TARE and ON/OFF buttons simultane-
ously to enter the setup menu.
 Wait until the first menu item < Cal> is displayed.
 Confirm by ➔ button, < CalExt> will be displayed.
 Use the navigation keys to select   < graadj>.
 Acknowledge using ➔ button, the current setting is dis-
played. The active digit is flashing.
 Enter weight value and confirm using the ➔ button, numer-
ical input see chap. see chap. 3.2.2.
Weighing balance returns to menu.
 Press repeatedly  button to exit menu.
25

Advertisement

Table of Contents
loading

Table of Contents