7.9.3 Gravitational constant adjustment location < graadj >
TFKB_B-BA-e-2210
Press and hold the TARE and ON/OFF buttons
simultaneously to enter the setup menu.
Wait until the first menu item <Cal > is displayed.
Confirm by ➔ button, < CalExt > will be displayed.
Use the navigation keys to select < graadj >.
Acknowledge using ➔ button, the current setting is
displayed. The active digit is flashing.
Enter weight value and confirm using the ➔ button,
numeric entry see chap. 3.2.2.
Weighing balance returns to menu.
Press repeatedly button to exit menu.