Safety Checks - Dräger Medical Vamos Technical Documentation Manual

Variable anesthetic gas monitor
Table of Contents

Advertisement

3.

Safety Checks

The following steps describe the safety
checks according to VDE 751 and
IEC 60601 (or UL 2601).
Perform electrical safety checks either
according to VDE 751 (chapter 22.3) or
IEC 60601 (chapter 22.4). The decision
whether to carry out safety checks according
to VDE 0751 or IEC 60601 should be based
on national regulations and/or regulations
made by the regions/countries, whichever is
applicable (for example, VDE 751 applies to
"Germany" region). Switching from one safety
check method to the other should be avoided.
Note: How to use the test devices "Secutest",
"GM50", and "NSE"/"Wison" is described in
the DORIS documentation (5000.4 MT Test
Equipment / Instructions for Use). The
"Wison" tester can only be used for testing
according to VDE 751.
3.1
Inspection
3.1.1
Power cable
3.1.2
Power pack with 15 V cable and connector.
Connect desktop power pack to VAMOS.
Switch on VAMOS.
3.2
Tests according to VDE 751
3.2.1
Protective earth test -not applicable-
(no protective earth connected)
3.2.2
Equivalent leakage current measurement - not
applicable - (no accessible metal housing
parts available)